![Larson Davis SoundTrack LxT Скачать руководство пользователя страница 207](http://html1.mh-extra.com/html/larson-davis/soundtrack-lxt/soundtrack-lxt_user-manual_3712247207.webp)
A-10
Technical Specifications LxT Manual
AC/DC Output
The purpose of the AC output is to drive a headset to listen to live and recorded sounds. It may be used
for other purposes, but may not function as expected as a source for additional analysis equipment.
The output is amplified for listening purposes and therefore is limited in its maximum output to be less
than the instrument’s maximum input level. For connection to external analysis equipment use the
adapter ADP015 and cable EXC006 to extract the signal directly from the preamplifier output. Using
the PRMLxT series preamplifiers puts a DC bias on the output of the BNC connector of the ADP015
of approxi 3.4 Volts.
AC Output Voltage Range
Vpeak maximum output
0.5 mV to 1.6 Vrms sine wave
(~70 dB dynamic range)
AC Output Gain
(relative to instrument input)
+ 39.2 dB
AC Output Frequency Weighting
Signal is unweighted with frequency limitations imposed by
hardware design (see below)
AC Output Frequency Response
20 Hz to 23.6 kHz (-3 dB), R
L
= 10 k
21 Hz to 23.6 kHz (-3 dB), R
L
= 600
95 Hz to 23.6 kHz (-3 dB), R
L
= 16
AC Output Recommended Loads
Headset with 16
or greater
speaker impedance
Resistive loads greater than 600
for maximum frequency
response range.
AC Output Impedance
Low impedance headset speaker driver with 100
F coupling
capacitor. Z = 1.5 + 1592/f, where Z is output impedance in
(Ohms) and f is frequency in Hz.
Instrument readings are not affected by AC output loading,
although with a large signal output, a short circuit may draw
excessive power and cause the instrument to turn off
.
AC Output Phase and Delay
- 180
relative to input, 128
s digital delay
DC Output Frequency Weighting
Follows the SLM Frequency Weighting: A, C or Z
DC Output Time Weighting
Follows the SLM Detector: F, S or I
Table A-9 AC/DC Output
Содержание SoundTrack LxT
Страница 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Страница 55: ...4 18 Measurement Settings Tabs LxT Manual...
Страница 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Страница 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Страница 110: ...LxT Manual Certification 7 17...
Страница 143: ...11 6 Control Panel Data Explorer LxT Manual...
Страница 167: ...12 24 Options LxT Manual...
Страница 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Страница 185: ...15 6 File System LxT Manual...
Страница 191: ...16 6 Miscellaneous Parameters LxT Manual...
Страница 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...
Страница 259: ...A 62 Technical Specifications LxT Manual...