HVM200 Reference Manual
Setting Up the Measurement
3-1
Section 3
Making Measurements
Selecting a Setup File by Using Your Mobile Device ............................................3-2
Positioning to Measure Hand/Arm Vibration ......................................................3-6
Positioning to Measure Whole Body Vibration with Seat Adapter .....................3-6
This module shows you how to perform a vibration measurement with the
HVM200 Control app and the G4 LD Utility.
3.1 Setting Up the Measurement
In this section:
• 3.1.1 Performing a Calibration Check
• 3.1.2 Selecting a Setup File by Using Your Mobile Device
3.1.1 Performing a Calibration Check
Before you begin:
• A calibration check requires a hand-held shaker, such as the PCB Model
394C06. To view shaker options, go to http://www.pcb.com/Sensor-
Calibration/Portable-Vibration
.
• Set sensitivity values for the x, y, and z axes on the
Sensor
• If needed, stop the measurement in progress.
Step 1.
Launch G4 or the HVM Control app on your mobile device.
Step 2.
Connect to your HVM200, then click the
Live View
. For help connecting,
see
2.6 Connecting HVM200 to G4 LD Utility
.
Step 3.
Click the menu icon
, then select
Cal. Check
in the menu that
appears.
FIGURE 3-1
Cal. Check Option
Содержание HVM200
Страница 1: ...Larson Davis HVM200 Reference Manual...
Страница 43: ...HVM200 Reference Manual A 8...
Страница 44: ...HVM200 Reference Manual A 9...
Страница 45: ...HVM200 Reference Manual A 10...
Страница 54: ...HVM200 Reference Manual A 19...
Страница 55: ...HVM200 Reference Manual A 20...
Страница 56: ...HVM200 Reference Manual A 21...
Страница 57: ...HVM200 Reference Manual A 22...
Страница 58: ...HVM200 Reference Manual A 23...
Страница 71: ...HVM200 Reference Manual B 4 B 1 7 ADP082A SEN041 FIGURE B 7 Frequency Response Function X Y and Z...
Страница 76: ...HVM200 Reference Manual E 1 Appendix E Declaration of Conformity...