1-10
Model 820 Components
Model 820 Technical Reference Manual
Block Diagram
All the standard functions of a precision sound level meter are
provided: instantaneous level, L
eq
, SEL, L
max
, L
min
, dose,
projected dose, etc. In addition, many valuable parameters can
be stored: time history interval data such as L
n
’s and L
eq
,
exceedance and exceedance time history, etc. Level calibration
is performed in a few key strokes, and every change or check
is entered in a calibration log.
The Model 820’s large data memory relieves the user from the
concern of data loss caused by memory limitations.
Furthermore, measurements can be printed or transmitted at up
to 19.2 K baud via RS-232 for further manipulation or
archiving.
Despite its numerous functions, the Model 820 is easy to use.
Annotated displays indicate units and measurement mode. The
single setup menu lets the user scroll through and program
parameters or use an index key to modify specific information.
Multiple measurement setups can be stored in memory for
later recall. And of course, complete setups can also be
transferred from a computer.
Figure 1-3
Block Diagram.
A
Filter
C
Filter
RMS 39
Slow, Fast, Impl
Peak 1
Peak 2
40
42
Weighted
Flat
Microphone
0/28/200 V 43
Bias Supply
41
A:D
Clock
Timer
LCD
Keypad
RAM
256K
Power
Supply
5-pin
Connector
Internal
9V Battery
AC/DC Output
20 dB
0 dB
Micro-
processor
Indicates parameter number
N
Flash
128K
820 Block Diagram