Preparation for NIBP monitoring :
1. Place the patient in supine position and connect the BP cuff to the arm.
2. If the cuff is not placed at level of the heart then the pressure values obtained
will not reflect the true physiological pressure.
Precautions with automatically cycled BP cuffs :
1. Place the limb in such a way to minimize stretching and avoid weight exertion on
affected nerves.
2 Select a measurement interval that provides adequate venous drainage during cuff
deflation.
3. The limb on which the cuff is connected must be inspected periodically in order to
detect venostasis.
4. An accurate BP determination might be difficult if the patient has irregular cardiac
rhythm.
5. Cuff size selection is a very important criteria to get accurate blood pressure readings.
6. Avoid compression or restriction of the NIBP pressure tube.
Note
Cuffs become soft after use. They sometimes develop folds which are permanent
and hence leave temporary marks on the limb. Any cuffs that exhibit this effect
should be replaced.
If NIBP is used with SpO
2
, (PR) pulse rate is derived from SpO
2
sensor.
If NIBP, SpO
2
and ECG all three are connected to patient; then heart rate (HR) is
derived from ECG.
Do not measure NIBP continuously, as it might result in prolonged impairment of the
blood circulation of the patient.
NON INVASIVE BLOOD PRESSURE
Note
Accuracy of NIBP performance / results depends on patients pre clinical condition and
the technique used (oscillometric method of calculating NIBP).
2.11
Содержание Star 55
Страница 1: ......
Страница 18: ...Star 55 1 1 DISPLAY CONTROLS CONNECTIONS AND OUTPUTS ...
Страница 19: ...Figure 1 1 23 15 19 16 17 22 18 21 20 1 234 5 6 7 8 9 10 11 12 13 14 Front Panel 1 1 ...
Страница 22: ...Star 55 Left blank intentionally 1 4 ...
Страница 23: ...Star 55 0843 1 2 Figure 1 2 Rear Panel 1 5 3 4 ...
Страница 25: ...Left Side Panel Figure 1 3 4 7 6 8 9 3 2 5 1 1 7 ...
Страница 27: ...Figure 1 5 1 Right Side Panel 2 3 1 9 ...
Страница 36: ...Star 55 2 APPLICATION GUIDE 2 ...
Страница 51: ...Star 55 Figure 2 15 2 15 ...
Страница 59: ...Star 55 OPERATIONS AND MAIN SCREEN 3 3 ...
Страница 61: ...3 2 MAIN SCREEN 1 Figure 3 1 Note Main screen when maximum i e eight waveforms are present ...
Страница 62: ...Star 55 MAIN SCREEN 2 Figure 3 2 Note Main screen when maximum ECG waveforms are present 3 3 ...
Страница 63: ...MAIN SCREEN 3 Figure 3 3 Note Main screen when single ECG waveform and three digital fields are present 3 4 ...
Страница 73: ...Star 55 4 PARAMETER SETTING 4 ...
Страница 91: ...Star 55 SYSTEM FUNCTIONS 5 5 ...
Страница 102: ...Star 55 Tachy High Figure 5 12 The Graphical Trend taken from the External Printer is as shown in the Figure 5 13 5 11 ...
Страница 103: ...Figure 5 13 5 12 ...
Страница 114: ...Star 55 6 MAINTENANCE SCHEDULE 6 ...
Страница 121: ...Star 55 ACCESSORIES 7 7 ...
Страница 130: ...Star 55 8 TROUBLE SHOOTING AND WAVEFORM 8 ...
Страница 150: ...Star 55 TECHNICAL SPECIFICATIONS 9 9 ...
Страница 166: ...Star 55 10 WARRANTY AND WARRANTY SERVICE 10 ...