Pulse Configuration
9.3
The pulse repetition frequency and the pulse voltage level that is fed into the attached field source
can be adjusted with the sliders depicted in Figure 22
Figure 22: Setting of the pulse repetition frequency and pulse voltage level
Figure 23: Pulse polarity
The polarity of the created pulses can be set in the GUI section
Polarity
. Available options are:
Positive (+)
Negative (-)
Setting of a Defined Pulse Counter
9.4
When enabled, the BPS 203 issues only the predefined number of pulses. After all pulses are
issued, the BPS 203 returns to the wait state.
Figure 24: Impulse counter
Probe Parameter Read Out
9.5
Depending on the attached field source, this field will show the resulting voltage / current /
electric field strength / magnetic flux density for a given pulse voltage level. These values are
hardcoded into the field
source’s memory during factory calibration.
When using field-coupled probes e.g. P1202 or P1301 the spacer ring used in the measurement
set-up has to be selected, since the electric field strength / magnetic flux density applied to the test
IC is highly dependant on the distance between field source and IC.
Figure 25: Probe parameter read out