5.9.2.6  Logging 

Data logging mode can be started or stopped via ‘Key 6 - Logging’ accessed from the 

‘Utilities’ menu. Press ‘Key 4 – Start/Stop Logging’ to start and stop data logging.  

 

Whilst in data logging mode the instrument will automatically record data at the pre-set 

intervals including running the pump for a preset time.  

 

Active logging mode is indicated on the ‘Main Read Screen’ by the   icon.   

 

The operator is able to edit the default

 

ID, pump

 

run

-

time, interval and start/stop 

logging. 

 

 

1)

 

From the ‘Main Read Screen’ press the ‘Menu’ key on the instrument panel. 

 

2)

 

Press ‘Key 6’ to display the ‘Logging’ menu and the following screen is displayed: 

 

 

 

 

 

 

 

 

 

 

 

 

Logging Menu 

 

 

3)

 

Select the desired option by pressing ‘Keys 1 to 4’. Then enter the appropriate 

setting using the keypad followed by the ‘Enter’ key.  

 

 

1-Every 00 mins. 

-  Press ‘Key 1’ to enter the time in minutes for the 

timeframe between sample readings.  The interval 

controls the reading frequency in minutes, i.e. every 10 

minutes.  

 

 

2-Pump 00 secs. 

-  Press ‘Key 2’ to enter the time in seconds for the length 

of time you wish the pump to run when taking a sample 

reading. The pump run-time is the time in seconds for 

which the pump runs prior to the reading being stored. 

This figure will also need to take into account the length 

of sample tube and the volume of the sample gas. For 

example, there is little point setting a pump run-time of 

10 seconds if it takes 30 seconds to draw in a new 

sample.  

 

3-ID 00000000 

-  Press ‘Key 3’ to create an 8-digit numeric ID Code. 

 

4-Start/Stop Logging  -  Press ‘Key 4’ to start and stop data logging. 

 

 

 Note:  Data logging mode is automatically stopped when the instrument is 

switched off or if the logging parameters are edited.  

 

 

Содержание Geotech G100 Series

Страница 1: ...rating Manual G100 CO2 0 20 G110 CO2 0 100 G150 CO2 0 10 000ppm Geotechnical Instruments UK Ltd Sovereign House Queensway Leamington Spa Warwickshire CV31 3JR England Tel 44 0 1926 338111 Fax 44 0 192...

Страница 2: ...OMG100N1 21 G100 Range Page 2 Copyright Geotechnical Instruments UK Limited This page is intentionally left blank...

Страница 3: ...nel 11 4 2 Panel Key Functions 12 4 3 Instrument Connection Points 13 5 0 GENERAL OPERATIONAL INSTRUCTIONS 14 5 1 Switching the Instrument On 14 5 2 Switching the Instrument Off 14 5 3 Instrument Main...

Страница 4: ...O2 Channel 33 7 5 5 Reset Factory Settings 33 7 6 Last Field Calibration 34 7 7 Calibration Record 34 8 0 PROBLEM SOLVING 35 8 1 Warnings and Errors 35 8 2 Flow Fail Warning 35 8 3 Self test Warning M...

Страница 5: ...ion 8 7 1 2 Safety Related Information Information in this manual that may affect the safety of users and others is preceded by the following symbol Warning Failure to follow this information may resu...

Страница 6: ...ification requirements which provide the user with a fast simple to use and accurate analyser Each model has been specifically designed to meet specific application requirements Apparatus intended for...

Страница 7: ...cal Research Laboratories IAQ Brewing Atmosphere Control Incubators Gas mixing 2 2 Instrument Components Standard Product G100 Analyser Reference A Analyser B Operating Manual C Mains Battery Charger...

Страница 8: ...can only be viewed via the optional Analyser Data Manager software It cannot be viewed on the analyser screen Applicable events are stored in the event log automatically No user intervention is requir...

Страница 9: ...itter or connection cable Make sure that the catches are aligned correctly Tighten the knurled nut by hand Note The humidity sensor can take 30 minutes to stabilise and special handling is required fo...

Страница 10: ...ead USBLEAD2 B Spare Sample Filters pack of 5 068296 S C Sample Tube Kit G1 6 D Hard Carry Case 050227 E F Temperature Probe 100mm Tip Temperature Probe 5mm Tip G1 3 G1 2 G H I Humidity Kit Probe 4mm...

Страница 11: ...Physical Characteristics of the Instrument Panel Front View Reference Back View Reference A Main Read Screen B Soft Keys C On Off Key D Pump Key E Key 4 Scroll Left F Key 8 Scroll Down G Menu Key H En...

Страница 12: ...nt panel are determined by menu options taken Functions vary from screen to screen C On Off Key Press the On Off key briefly to switch the instrument on and off D Pump Key Press the Pump key to start...

Страница 13: ...emperatures Temperature connectors are fitted with black plugs to prevent dust ingress remove before use Grip connector cap and pull upwards to release Refit when not in use C Humidity Probe Optional...

Страница 14: ...hed off a clean air purge should be performed This ensures that the instrument is free from gas and ready for the next measurement This final purge is especially important for the oxygen sensor as it...

Страница 15: ...connected Humidity probe connected flashing Service due every 12 months Service overdue Fault repair Waiting Note A red backlight is displayed if the pump is stalled or the alarm is activated For fur...

Страница 16: ...ty Continue to press the Scroll key to return to the Main Read Screen 5 7 Memory The memory should not be used as a permanent storage medium and any important data should be transferred to a more perm...

Страница 17: ...on Screen 5 9 2 Utilities The Utilities option enables the operator to configure instrument settings prior to taking readings 1 From the Main Read Screen press the Menu key on the instrument panel 2 P...

Страница 18: ...The Contrast option enables the operator to adjust the instrument screen contrast to compensate for changes in ambient temperature The default setting is 0 1 From the Main Read Screen press the Menu...

Страница 19: ...when the gas level falls below the minimum value entered by the user 1 From the Main Read Screen press the Menu key on the instrument panel 2 Press Key 2 to display the Utilities menu 3 Press Key 3 to...

Страница 20: ...The Flow Fail option enables the operator to adjust the instrument flow fail detection point should it fail in normal operation with a clean filter The instrument s internal pump can be stalled when...

Страница 21: ...ail alarm Flow fail alarm No Wait 5 seconds No Block gas inlet Flow fail alarm Wait 5 seconds No Block gas inlet Flow fail alarm Wait 5 seconds Accept Blocked or excessive restriction No Wait 5 second...

Страница 22: ...o 4 Then enter the appropriate setting using the keypad followed by the Enter key 1 Every 00 mins Press Key 1 to enter the time in minutes for the timeframe between sample readings The interval contro...

Страница 23: ...can be performed Note For further information please refer to section 7 0 Calibration of this manual 5 9 4 View Data The View Data option enables the operator to view the stored readings 1 From the Ma...

Страница 24: ...ng is activated the instrument will show a brief message stating that the memory is full and that no further data will be recorded 1 From the Main Read Screen press the Menu key on the instrument pane...

Страница 25: ...nt will power on and display charging When complete the display will change to show that the instrument is charged To switch the instrument ON whilst charging or charged is displayed the operator will...

Страница 26: ...6 If necessary check the span calibration with a known concentration calibration check gas Step 7 Take readings Do protect the instrument from strong direct sunlight which will quickly raise the temp...

Страница 27: ...ings have stabilised approx 30 seconds then press the soft key Store 5 The pump will stop and the operator will be prompted to enter an ID code to identify the reading A reading stored confirmation me...

Страница 28: ...ading interval and pump run time These parameters are used to control the reading frequency in logging mode Once logging mode is activated the instrument will automatically record a reading at every i...

Страница 29: ...ly displayed reading This allows it to be manually recorded or moved away from the sample point Once activated press the soft key Hold and the readings are fixed until the Hold key is pressed again or...

Страница 30: ...range Users should select the correct calibration gas for the expected gas levels on their particular application Only use gases with a known certified gas concentration Note Certified calibration gas...

Страница 31: ...regulator supplied with the calibration kit is recommended as flow and pressure rates are factory set Note Maximum input pressure 250mb maximum flow 300ml 7 5 Calibration Method Before you begin ensur...

Страница 32: ...from the menu Then either attach the 100 N2 or sample pipe to allow access to fresh air 3 Ensure the zero gas has flowed and is stable 4 Press the Start key The instrument will now wait approximately...

Страница 33: ...concentration of your calibration gas 2 Press the soft key Start and wait for the reading to stabilise Press the Pump key to draw in fresh air It can take a couple of minutes to stabilise 3 Once a st...

Страница 34: ...record the following in the event log For each entry the time and date will be stored Event Data Recorded Successful user zero CO2 Type N2 or Air and Readings before and after Successful user span CO...

Страница 35: ...ser calibration or may indicate sensor failure If an incorrect user calibration has caused the warning it should be correctable by way of returning the instrument to factory settings zeroing or carryi...

Страница 36: ...al This warning is most commonly caused by an incorrect user calibration Try recalibrating the sensor or press Return to factory settings If the warning persists it may be caused by dirt or damage to...

Страница 37: ...ated The instrument will need to be returned to the manufacturer for service Note Certain configuration problems can be corrected remotely Using the Analyser Data Manager software it is possible to ex...

Страница 38: ...entire curve for lamp and filter variations caused by aging and user induced drift due to dirt etc If done correctly there is often no need to complete a span calibration However a poor calibration A...

Страница 39: ...2 SO2 or H2 unlike many other types of oxygen cell 8 6 Error Due to CO2 Solubility in Water Due to the water trap and filter it is possible that some of the CO2 in the sample gas will be absorbed in t...

Страница 40: ...an be user serviced Sample Filter This should be regularly inspected for damage or discolouration and changed if needed The instrument should never be operated without the sample filter as this may re...

Страница 41: ...period Geotech UK Limited will repair or replace defective parts on an exchange basis The decision to repair or replace will be determined by Geotech UK Limited To maintain this warranty the purchaser...

Страница 42: ...ype 0 Readings 1 Event Log Change contrast Before and After Change flow fail current limit Before and After Battery less than critical voltage Critical Actual RTC date time invalid None Factory calibr...

Страница 43: ...1 21 Copyright Geotechnical Instruments UK Limited Page 43 12 0 Sample Certificate of Calibration This is a sample certificate of calibration supplied at the time of purchase and updated when the inst...

Страница 44: ...s producers responsible from July 1st 2007 in meeting their obligations with the fundamental aim of reducing the environmental impact of electrical and electronic equipment at the end of its life Geot...

Страница 45: ...ove is are in compliance with the following standards EMC Directive 2004 108 EC EN 50270 2006 Electromagnetic compatibility Electrical apparatus for the detection and measurement of combustible gases...

Страница 46: ...d to clear out gas from the inlet pipe and the analyser s gas sensors prior to taking a new reading CO2 Carbon dioxide gas Contrast adjustment Adjustable setting which darkens or lightens the text dis...

Страница 47: ...other applications such as MS Excel General warnings Displayed throughout the documentation with a warning symbol Warning information may affect the safety of users Inlet port Port located on the top...

Страница 48: ...present Temperature probe External device to enable the instrument to display data and record the temperature readings This is an optional feature Under range codes Under range codes are errors below...

Отзывы: