GEM5000 Gas Analyzer
OMGEM5KN4.7.1
Page
23
of
70
7.2.2 Gas Check
This option displays the ‘Gas Check’ menu and enables the operator to zero
and span the gas channels on the instrument. Historical/previous gas
checks data can also be viewed and factory settings can be restored.
1)
Select the ‘Menu’ button on the front of the analyzer to display the ‘Device Settings’
menu. Press the soft key to display ‘Operation Settings’.
2)
Select ‘Key 2 – Gas Check’ and the following menu is displayed:
3)
For more information about the Gas Check Menu please refer to section 9.0 –
Calibration.
4)
Select soft-key ‘Exit’ to exit operation settings and return to the main screen.
7.2.3 View data
This option enables the operator to view the readings collected and stored
on the instrument. Readings many be downloaded to the optional LSGAM
software if further analysis is required.
1)
Select the ‘Menu’ button on the front of the analyzer to display the ‘Device Settings’
menu. Press the soft-key to display ‘Operation Settings’.
2)
Select ‘Key 3 – View Data’ and the following screen is displayed: