Page 6

User Guide

198.240

3.0

SERVICES

3.1

Electrical Connections

The LSP Linescanning Thermometer requires just one electrical interface for normal operation. This interface
is made to the LSP Processor, via the interface cable (Land part no. 031.402) supplied with the system. The cable
is supplied pre-terminated at the scanner connection end, but the processor end is unterminated to allow the
cable to be cut to suit the application. The plug at the scanner end of the interface cable requires care when
making the connection to the scanner socket. Simply orientate the plug to suit the socket and press into place.
Do not force the plug into place. To remove the plug, grip the knurled sleeve section of the plug and pull upwards.
Termination details for the processor end of the interface cable can be found in the Land LSP Processor
Installation Guide.

Caution

The interface cable may be cut to length, but under no circumstances should an attempt be made to
extend the cable length beyond the 15m / 49.2ft maximum, by joining to a second length of cable. Failure
to comply with this instruction could result in incorrect operation of the instrument.

3.2

Water Requirements (Water-Cooled Units Only)

When the LSP Linescanning Thermometer is to be used in an environment where ambient temperatures are
above the maximum specified, the Land water cooled, air purge mounting accessory must be used. This requires
a constant water supply to be available. The special mounting plate has internal water channels with inflow and
outflow connections (G1/8 tappings) in the plate edge. There is no preferred flow orientation, however, if the
scanner is mounted in a position where one of the water connectors is higher that the other, the lower connector
must be used for the mains water in-flow.

The water-cooled mounting plate is supplied fitted with hose connectors suitable for 9.5mm / 3/8in bore water
hose. The water flow rate required will depend upon each individual application, but 1l/min / 0.26USgal/min is
an absolute minimum at 30°C / 86°F maximum water temperature.

Caution

Overcooling of the unit to below the local dewpoint, indicated by the formation of condensation, must
always be avoided. The temperature of the cooling water should ideally be in the range of
15 to 30°C / 59 to 86°F.

A clean, constant water supply is recommended, as this reduces the likelyhood of blockages within the
mounting plate.

Содержание LSP50

Страница 1: ...inst com Internet www landinst com Land Instruments International 2003 Publication N LSPS198 240 Issue A 05 03 Land Instruments International 10 Friends Lane Newtown PA 18940 1804 U S A Telephone 215...

Страница 2: ......

Страница 3: ...agement System of Land Instruments International Ltd is approved to BS EN ISO9001 2000 for the design and manufacture stockholding in house repair and site servicing of non contact temperature measuri...

Страница 4: ...current IEC 617 2 No 02 02 06 Three phase alternating Current IEC 417 No 5017 Earth ground terminal IEC 417 No 5019 Protective conductor terminal IEC 417 No 5020 Frame or chassis terminal IEC 417 No 5...

Страница 5: ...SO 3864 No B 3 6 Caution ISO 3864 No B 3 1 WARNING Risk of electric shock BS EN 100015 Observe precautions for handling electrostatic discharge sensitive devices Refer to the operating instructions BS...

Страница 6: ...Blank User Guide 198 240...

Страница 7: ...allation Precautions 3 0 SERVICES 5 3 1 Electrical Connections 3 2 Water Requirements Water Cooled Units Only 3 3 Air Requirements Air Purged Cooled Units Only 4 0 LASER ALIGNMENT 7 4 1 Using The Lase...

Страница 8: ...1 Introduction 8 0 MAINTENANCE 14 8 1 Air Purge Services 8 2 Water And Air Cooling Services 9 0 INSTALLATION ACCESSORIES 15 9 1 Basic Mounting Plate 9 2 Right Angle Mounting Bracket 9 3 Water Cooled...

Страница 9: ...n operation and maintenance of the LSP Linescanning Thermometer and should be used in conjunction with the information detailed within the LSP Linescanning Thermometer Installation Guide LSP Processor...

Страница 10: ...Blank User Guide 198 240...

Страница 11: ...SYSTEM comprises LSP Linescanning Thermometer LSP Processor LSP Interconnection Cable LANDSCAN NT Software Optional LANDSCAN for Windows Configuration Software Standard supplied with LSP Processor or...

Страница 12: ...el displays the scanner serial number the model number and the CE conformity mark The model number indicates specific information about the unit including Wavelength and Temperature Range Example Syst...

Страница 13: ...100Hz in 10Hz steps TargetWidth Targetdistance 1200mm 47 2in 12mm 0 5in Targetdistance 1200 to 3000mm 47 2 to 118 1in FoV 100 1 Speed Of Response Tau on rising edge linearoutput 7 s AmbientTemperatur...

Страница 14: ...capable of carefully following the procedures warnings and guidelines given It is adviseable that all User Guides and Installation Guides associated with LSP System equipment should be read thoroughly...

Страница 15: ...e scanner mountings ensure that the unit is replaced in the correct orientation should the scanner be removed from the application for servicing etc 2 4 Installation Precautions Vibration It is highly...

Страница 16: ...could result in incorrect operation of the instrument 3 2 Water Requirements Water Cooled Units Only When the LSP Linescanning Thermometer is to be used in an environment where ambient temperatures ar...

Страница 17: ...mounting plate If possible ensure that the air purge exit direction is downwards and ensure that this exit is not impeded by mounting structures The air purge flow rate depends upon local conditions...

Страница 18: ...arning CLASS 2 Laser Product DO NOT attempt to disassemble the laser unit or any of it s mounting components Embedded laser has a maximum output of 3mW with a beam divergence of 5mrad Caution Caution...

Страница 19: ...laser light is projected through the window of the scanner indicating the position and extents of the scan line Adjust the scanner position as required to ensure that the product almost fills the sca...

Страница 20: ...Monel oxidised 0 45 Glass 0 90 3mm Nichrome oxidised 0 90 Paper 0 90 Polypropylene METALS Polyethylene 0 90 2mm thickness Aluminium oxidised 0 30 0 96 5mm Copper oxidised 0 85 Iron Steel oxidised 0 85...

Отзывы: