WARNING
The operator shall not touch the ther
-
mometer and the patient simultaneously.
Keep the machine out of children’s reach.
* For accidental swallow of bat
-
tery or protective film, please con
-
sult the doctor at once.
For temperature difference between stor
-
age area and measurement site, condi
-
tion the thermometer for about 30 minutes
in room temperature (measurement site).
* Failing to do so may cause inaccuracy.
Ensure that children do not use
the instrument unsupervised.
* Some parts are small enough
to be swallowed.
ERROR MESSAGE
ERROR
MESSAGE
PROBLEM
SOLUTION
Hi
Tempera-
ture taken
is higher
than 42.2°C
(108°F).
Operate the ther-
mometer only between
the specified temper
-
ature ranges. In the
event of a repeated
error message, con
-
tact your retailer
or Customer Services
Lo
Tempera-
ture taken
is lower
than 32°C
(89.6°F).
Operate the ther-
mometer only between
the specified temper
-
ature ranges. In the
event of a repeated
error message, con
-
tact your retailer
or Customer Services
Err
Sensor not
welded
Contact your
retailer or Cus
-
tomer Services
ErE
EEPROM is
abnormal
Contact your
retailer or Cus
-
tomer Services
Low power
Replace the
batteries
ErH
Temperature
of the
measuring
environment
is too high.
Decrease the
temperature of the
environment, keep
it within 16°C-35°C.
ErL
Temperature
of the
measuring
environment
is too low.
Increase the
temperature of the
environment, keep
it within 16°C-35°C.
MAINTENANCE, STORAGE
AND CALIBRATION
Maintenance
1
Remove any stains on the body
with a soft and dry cloth.
2
Gently wipe the dirt on the body with a soft cloth,
use a cotton swab or soft cloth stained with alcohol,
and gently wipe the lens of the temperature mea-
suring head.After using the product, the product
can be sterilized with 75% medical alcohol on the
soft towel or cotton sliver.
If the probe tip is damaged, please
contact the maintenance center.
3
Don’t wash with water or detergent
containing abrasive or benzene.
4
Do not immerse in liquids.
Storage
1
The device must not be stored or used at an exces-
sively high or low temperature or humidity, in sun-
light, in association with an electrical current or in
dusty locations. Otherwise may occur inaccuracies.
2
Remove the battery if non-use for long time.
Calibration
This thermometer is calibrated at the time of manufactured.
If the thermometer is used according to the instruction,
periodic recalibration is not required. If at any time you
question the accuracy of the measurement, please contact
the retailer immediately. Don’t attempt to modify or
reassemble the thermometer.
PARAMETER SPECIFICATION
Measure
-
ment Mode
Forehead mode
Measur
-
ing site
Forehead
Measure
-
ment Units
Celsius (°C) and Fahr
-
enheit (°F)
Resolution
0.1 (°C/°F)
Operating
Conditions
16°C~35°C (60.8°F~95°F) with a
relative humidity of 15%~80%
Storage Con-
ditions
-20°C~+55°C (-4°F~131°F) with a
relative humidity of 15%~93%
Measurement
distance
15mm to 50mm from the
measuring point
Forehead
temperature
measurement
range and
accuracy
Measurement range:
32.0°C~42.2°C (89.6°F~108°F)
Measurement accuracy (Labra
-
tory) ±0.3°C (32.0~-42.2°C)
/ ±0.5°F (89.6°F~108°F)
Memories
recall
32 measurements recall
Dimensions
138mm
×
37mm
×
37mm
Weight
47g without batteries
Power
2
×
AAA batteries (DC 3V)
Switch off
Automatically switches
off after 1 minute
Product Life 5 years
New bat-
tery life
1000 times
STANDARD
The product is made under the ISO80601-2-56.
GUARANTEE
Two-year warranty is available from purchasing date,
excluding user-caused failures listed below: