Maintenance
5.1.8
Premature Cold Start Termination
Occasionally, it might be necessary to terminate the COLD START prematurely, e.g., after a
short-term power failure, but not before a minimum waiting period of 120 seconds has expired.
NOTICE
A premature cold start termination leads directly into measurement operation without
an additional prompt. If the probe does not reach its optimum operating point, falsified
measured values and, possibly, faults and warnings are the result.
In the main menu, select
for the calibra-
tion of the probe.
Interrupt the cold start prematurely.
By selecting with the cursor key
and
ENTER
the cold start is terminated pre-
maturely.
The system returns to the main menu auto-
matically.