10.2 Technical data of the KS1-DK combination probe
Measuring range:
O
2
:
0...18 vol. % O2
with limitation 0...21 vol. % O
2
COe: 0...1000ppm COe
Measurement accuracy:
O2:
±
10 % of measurement value
No better than
±
0.3 vol. % O
2
CO:
±
25 % of measurement value
No better than +/- 10 ppm in exhaust gases from natural gas
combustion following previous calibration under operating
conditions with a CO reference measurement in the measuring
range 0...100 ppm :. 10 ppm
Probe output voltage:
O
2
electrode:
-100mV...+1000mV
Resolution:
0.1mV
CO/H
2
electrode:
-50...+950mV
Resolution:
1mV
Error-influencing factors:
Temperature
Other unburned hydrocarbons
Cross-sensitivity:
To SO
2
, NH
3
, NO, propane, aromatic hydrocarbons
Probe internal resistance of ZrO
2
measuring cell in air and 22W
heating power:
15...25 ohms
Permitted fuels:
Residue-free, gaseous hydrocarbons
Light fuel oil
Coal
Wood
Permitted continuous exhaust gas
temperature:
< 450 °C
Life:
≥
3 years with natural gas
Heating power:
20...25 watts, depending on the version and the measuring gas
temperature
Heating current:
Approx. 1.3 A
Insulation resistance between
heater and probe connection:
> 30 M
Ω
Mounting position:
Horizontal to vertical
Protection class:
IP42
Counter flange:
In version standard
Æ
DN65
In version blow.out device
Æ
DN85