Lampert DENTA PUK Скачать руководство пользователя страница 14

 

14 

If then the foot switch is pressed, the welding process will be triggered.  

Please note!

  No gas is released during the entire pin welding process. 

 

 

Please note! 

During step “2.” Above (whilst the shutter is flickering), if the foot switch is not pressed 
within 5 seconds, a constant signal tone will commence. This is a warning, and is 
designed to protect the hand-piece solenoid from overloading in the case of unintentional 
contact. 

 

 

By again holding the footswitch pressed for ca. 1 second, (without contacting the work 
piece) the footswitch function can be deactivated. The symbol 

 will disappear from the 

display. 

 

 

 

6-3  WORKING WITH THE ADAPTER CABLE FOR WELDING PINS (Ø1,3mm)  

 

  

 
The pin welding mode gives the user the possibility weld on pins (Ø 1,3 mm). Here the 
provided adapter cable is utilised, to enable a better contact to the work piece.  
 
After the pin is welded on to the work piece, the cable can simply be slotted on to the pin, 
thus providing the optimum contact with the DENTA PUK, so that further welding work 
can be carried out. (This is especially advantageous for all work where contacting the 
work piece with the other provided tools (i.e. crocodile clip), proves difficult. 

The pin is only a “temporary aid” that must not withstand high levels of 
mechanical stress. As it will be removed again after the welding work is 
completed, it is important to remember to use a low power setting when welding it 
on. This will ensure that the pin can be easily removed and polished afterwards. 

     

 

  

This can be a very useful method, i.e. when wanting to carry out a weld on a model cast 
denture that is already mounted to a model.  

 

  

It is best to choose a place on the work piece that is sufficient in size and is not prone 
warping or distortion; also important, is that the chosen spot for the pin is easy to finish 
and polish afterwards.    

 

 

After the work is finished, the pin can be easily removed again, and the welding area 
cleaned with the rest of the work piece after all welding is completed.  

 

 

6-4  TIPS AND GENERAL POINTERS 

 

  

Important ! 

 

  

Always work with a well sharpened electrode !  

 

  

This is the only way to achieve optimum results. 

 

  

Always make sure that the work piece has a good contact to the 
connecting terminal (i.e. crocodile clip).  
In case of problems that arise as a result of poor electrical contact, 
attach the connecting terminal to a part of the work piece where the 
bare metal is exposed. 

 

  

Never weld “freehand“, meaning: always lay both hands on the 
hand-rests; this will aid in steadying the hands. If the hands shake, 
the parameters of the device can be falsified and the welding result 
affected.  

 

  

Do not apply force when the electrode tip touches the work piece; 
work using no pressure, or only the slightest of pressure ! 

 

  

 

  

Weld using only a low gas pressure ! 
In most cases about 2 l/min is enough.  

Содержание DENTA PUK

Страница 1: ...ration of the machine Before first using your DENTA PUK please be sure to carefully read the manual sections General Safety Requirements and Personal safety Please retain these instructions for future...

Страница 2: ...NG PRINCIPLES AND SETTING THE PARAMETERS 9 5 1 Adjusting the welding parameters 10 5 2 Using the programmable memory 11 6 INSTRUCTIONS FOR USE 6 1 Welding instructions 12 6 2 Welding with the foot swi...

Страница 3: ...y Any other application of the appliance other than the above stated is prohibited It is prohibited to use this apparatus out of doors Use only in dry surroundings No liability of any kind will be ass...

Страница 4: ...ential for electrical hazard of the workplace If it can be assumed that a safe operation of the device is no longer possible the machine must be shut down and removed from the power supply it must als...

Страница 5: ...e Those lingering or working in close proximity to the electric arc must be made aware of the risks and supplied with the appropriate protective equipment Wherever necessary a protective screen should...

Страница 6: ...m indicator P1 P10 P1 P5 are pre programmed at the factory The foot switch function can be recognised by this symbol The foot switch itself is an optional accessory Please note In the following cases...

Страница 7: ...lare protection system and the LED microscope lighting are identical 16 CONNECTOR FOR INERT GAS HOOKUP ARGON GAS For pressure hose 6 mm 17 IDENTIFICATION PLATE 18 FUSE TRAY Fuse 19 SERIAL NUMBER 4 4 S...

Страница 8: ...the DENTA PUK Other glare protection systems are not permitted and can result in permanent health damage or lead to the damaging of the machine Please note Before starting to weld always check to make...

Страница 9: ...le clip and cables receive electrical current Please see to it that these items cannot come into contact with any electrically conducting or earthed materials For the user operator there is no danger...

Страница 10: ...e in the machine will now be open and the gas flowing freely Without letting go of the rotary knob welding power 4 adjust the flowmeter regulator to the correct flow rate ca 2 litres per minute 5 1 SE...

Страница 11: ...ues for normal material thicknesses Especially when working with thin or delicate constructions it is advisable to start with lower output levels so that the work piece doesn t become damaged if neces...

Страница 12: ...the electrode Maintain the contact until the weld has been produced Please note Do not apply force when the electrode tip touches the work piece work using no pressure or only the slightest of pressu...

Страница 13: ...ol In Standard Pulse Soft Micro mode Attach one of the terminal clamps i e crocodile clip to a part of the work piece where the bare metal is exposed If the work piece is brought into contact with the...

Страница 14: ...n after the welding work is completed it is important to remember to use a low power setting when welding it on This will ensure that the pin can be easily removed and polished afterwards This can be...

Страница 15: ...n be helpful use an appropriate welding wire 6 5 SHARPENING THE ELECTRODES Please switch the machine off before changing the electrode this is a safety precaution which will safeguard against the unco...

Страница 16: ...Maximum Gas pressure 4 bar Protection category I Insulation class B Degree of protection IP 21S Weight DENTA PUK 8 8 kg 7 2 Identification plate Explanation of picture symbols Amperage Voltage Degree...

Страница 17: ...y fuse the mains power supply Mains fuse blows as soon as the machine is switched on The machine will need servicing please contact your dealer 5 Bad welding result bad welding characteristics The wro...

Страница 18: ...h care they undergo meticulous testing processes to ensure high quality and length of product life Nevertheless should the welding device malfunction you are assured of our support and a competent ser...

Страница 19: ...k GmbH Ettlebener Str 27 D 97440 Werneck Declares herewith that the following product Precision welding device DENTA PUK Complies with the provisions of the below mentioned directive including any ame...

Отзывы: