11
Unstable readings
Insufficient electrolyte
in probe. Air bubbles
visible when probe
inverted.
Electrolyte is depleted
Replace electrolyte, and
membrane cap assembly
Readings drift down Insufficient
stirring
Stir sample with meter or move
probe in sample
Slow response
Dirty or damaged
membrane
Replace electrolyte and
membrane cap
Electrode can not be
calibrated
Depleted electrolyte
Dirty or damaged
membrane
Replace electrolyte and
membrane cap
Electrode can not
be calibrated after
replacing electrolyte
and membrane cap
Dirty probe.
Cathode is not a shiny
gold color.
Clean cathode with cleaning
cloth or cleaning paper. Do not
over polish.
Sample reading is
frozen
Meter is in HOLD mode
Meter is locked
Release HOLD by momentarilly
pressing
MODE
button.
Remove batteries. Press
ON/
OFF
button. Replace batteries
and press
ON/OFF
button.
Stored readings will be lost.
MAINTENANCE
Battery Replacement
1. Twist off the battery compartment cover.
2. Holding the battery housing in place with a finger, pull out the
battery carrier using the two small tabs.
3. Replace the four (4) CR2032 batteries observing proper
polarity.
4. Replace the battery carrier, reattach the battery
compartment cap and tighten securely.
Battery Safety
¬
Dispose of batteries responsibly; observe local, state,
and federal regulations.
¬
Never dispose of batteries in a fire; batteries may explode or leak.
¬
Never mix battery types; install new batteries of the same type.
Electrode Replacement
1. Turn the meter OFF. Turn the electrode retaining collar clock-wise and
remove it.