9
NOTE: The meter will remember the last scanned blank reading. It is not
necessary to scan a blank each time the test is performed. To use the
previous blank reading, instead of scanning a new one, scroll to Scan Sample
and proceed. For the most accurate results, the meter should be blanked
before each test and the same tube should be used for the blank and the
reacted sample.
NOTE: For the best possible results, carry a reagent blank through the
procedure. After scanning the blank, perform the test procedure on clear,
colorless, distilled or deioized water. Subtract results of regent blank from all
subsequent test results.
CALIBRATION
STANDARDS
The meter should be calibrated with sulfate standards. The calibration should
be done with a distilled or deionized water blank and one sulfate standard of
known concentration. The concentration of the calibration standard should be
similar to the expected concentration of the sample that will be tested.
CALIBRATION PROCEDURE
1. Press and briefly hold
to
turn the meter on. The LaMotte
logo screen will appear for about
3 seconds and the
Main Menu
will appear.
Main Menu
Testing Menu
Editing Menu
12:00:00 001/500
2. Press
ENTER
to select
Testing
Menu
.
All Tests
089 Sulfate HR
12:00:00 001/500
3. Scroll to and select
089 Sulfate
HR
from menu.
089 Sulfate HR
Scan Blank
Scan Sample
12:00:00 001/500
Calibration