9. Insert tube into chamber. Close
lid. Select
Scan Sample
.
089 Sulfate HR
9
ppm
Scan BLank
Scan Sample
12:00:00 001/500
10. Press
to scroll to
Calibrate
.
089 Sulfate HR
9
ppm
%T/Abs
Calibrate
12:00:00 001/500
11. Press
ENTER
to select
Calibrate
.
A reverse font (light background
with dark characters) will appear
to indicate that the reading can
be adjusted.
089 Sulfate HR
9
ppm
∧
,
∨
=Edit, ENTER=Save
∧
+ENTER=Default
12:00:00 001/500
12. Press
or
to
adjust the value shown to the
concentration of the prepared
standard, 10 in this example.
NOTE: A maximum adjustment
of 25% is possible.
089 Sulfate HR
10
ppm
∧
,
∨
=Edit, ENTER=Save
∧
+ENTER=Default
12:00:00 001/500
Calibration