7
ANALYSIS
PROCEDURE
1. Fill the Water Sample Collecting
Bottle (0688) with sample
water.
2. Press and hold
until
colorimeter turns on.
Main Menu
Testing Menu
Editing Menu
12:00:00 001/500
3. Press
ENTER
to select
Testing
Menu
.
All Tests
081 Phosphate LR
12:00:00 001/500
4. Scroll to and select
081
Phosphate LR
from menu.
081 Phosphate LR
Scan Blank
Scan Sample
12:00:00 001/500
5. Rinse a clean tube (0290) with
sample water. Fill to the 10 mL
line with sample. Cap and wipe
dry.
Analysis