The display may show the following messages:

Battery Status

 

More choices are available and can be viewed by scrolling 
up and/or down through the display.

Header

Identifi es the current menu and information on units and 
reagent systems if applicable. 

Footer

In the data logging mode the number of the data point 
is displayed and the total number of data points in the 
memory will be shown. The footer also shows current time 
and battery status

 

„

TUBES AND CHAMBERS

The colorimeter uses one type of tube (Code 0290) for all test factors. 

The handling of the tubes is of utmost importance. Tubes must be clean and 
free from lint, fi ngerprints, dried spills and signifi cant scratches, especially 
the central zone between the bottom and the sample line.

Scratches, fi ngerprints and water droplets on the tube can cause stray light 
interference leading to inaccurate results. Tubes that have been scratched in 
the light zone through excessive use should be discarded and replaced with 
new ones.

Tubes should always be washed on the inside and outside with mild detergent 
prior to use to remove dirt or fi ngerprints. The tubes should be allowed to air-
dry in an inverted position to prevent dust from entering the tubes. Dry tubes 
should be stored with the caps on to prevent contamination.

After a tube has been fi lled and capped, it should be held by the cap and the 
outside surface should be wiped with a clean, lint-free absorbent cloth until it 
is dry and smudge-free. Handling the tube only by the cap will avoid problems 
from fi ngerprints. Always set the clean tube aside on a clean surface that will 
not contaminate the tube. It is imperative that the tubes and light chamber be 
clean and dry. The outside of the tubes should be dried with a clean, lint-free 
cloth or disposable wipe before they are placed in the meter chamber.

Tubes should be emptied and cleaned as soon as possible after reading a 
sample to prevent deposition of particulates on the inside of the tubes. 

Variability in the geometry of the glassware and technique is the predominate 
cause of variability in results. Slight variations in wall thickness and the 
diameter of the tubes may lead to slight variations in the test results. To 
eliminate this error the tubes should be placed in the chamber with the same 
orientation each time.

www.

.com

[email protected]

1.800.561.8187

Содержание DC1500 PHOSPHATE

Страница 1: ...DC1500 PHOSPHATE Phosphate Colorimeter Kit Code 3242 www com information itm com 1 800 561 8187...

Страница 2: ...Data 14 Factory Setup 15 Setting Power Save 15 Setting Backlight Time 16 Selecting A Language 17 Looping Menus 18 Computer Connection Output 19 Computer Connection 19 WaterLink Connect 19 Battery Bat...

Страница 3: ...25 General Precautions 25 Safety Precautions 26 Limits of Liability 26 Speci cations 27 CE 28 IP67 28 Warranty 28 Register Your Meter 28 Trouble Shooting Guide Error Messages 29 Calibration 29 Stray L...

Страница 4: ...PH 1 USB Wall Adapter 1721 1 USB Cable 1720 01 1 DC1500 PH Phosphate Colorimeter Manual 3242 MN 1 DC1500 PH Phosphate Colorimeter Quick Start Guide 3242 QG To order individual reagents or test kit com...

Страница 5: ...to an intense blue colored complex by ascorbic acid The color is proportional to the amount of phosphate present Only orthophosphate forms a blue color in this test Polyphosphates and some organic pho...

Страница 6: ...g Menu 12 00 00 001 500 3 Press ENTER to select Testing Menu All Tests 081 Phosphate LR 12 00 00 001 500 4 Scroll to and select 081 Phosphate LR from menu 081 Phosphate LR Scan Blank Scan Sample 12 00...

Страница 7: ...l turn blue if phosphates are present Wip tube dry 05 00 10 Insert tube into chamber Close lid Select Scan Sample Record result as Ammonia nitrogen 081 Phosphate LR 0 99 ppm Scan BLank Scan Sample 12...

Страница 8: ...tilled or deionized water blank and one phosphate standard of known concentration The concentration of the calibration standard should be similar to the expected concentration of the sample that will...

Страница 9: ...ct Scan Blank 6 Remove tube from colorimeter Use 1 0 mL pipet 0354 to add 1 0 mL of Phosphate Acid Reagent V 6282 Cap and mix 7 Use the 1 0 g spoon 0699 to add one measure of Phosphate Reducing Reagen...

Страница 10: ...verse font light background with dark characters will appear to indicate that the reading can be adjusted 081 Phosphate LR 0 99 ppm Edit ENTER Save ENTER Default 12 00 00 001 500 12 Press or to adjust...

Страница 11: ...ent press EXIT Press and ENTER at any time to return to the default value The calibration has now been standardized and can be used for testing Scroll to Scan Blank and begin testing 081 Phosphate LR...

Страница 12: ...select Set Clock The year is displayed Press or to scroll to the appropriate character Press ENTER to select the character The month day hour format hour minute second AM PM will be dislpayed Repeat...

Страница 13: ...scroll to Logging Editing Menu Set Clock Logging Factory Setup Set Power Save 12 00 00 001 500 2 Press ENTER to select Logging Logging Display Test Log Logging Enabled Logging Disabled Erase Log 12 0...

Страница 14: ...to scroll to Set Power Save Editing Menu Set Clock Logging Factory Setup Set Power Save 12 00 00 001 500 2 Press ENTER to select Set Power Save Set Power Save Disable 5 Minutes 15 Minutes 30 Minutes 1...

Страница 15: ...more frequently the battery will have to be charged if the USB Wall Adapter is not being used 1 From the Editing Menu press or to scroll to Backlight Time Editing Menu Logging Factory Setup Set Power...

Страница 16: ...o scroll to desired language Select Language English Spanish French Portugese 12 00 00 001 500 4 Press ENTER The screen will display Storing for about 1 second and the meter will return to the Editing...

Страница 17: ...ion The diagrams below demonstrate a looping menu AND SO ON AND SO ON AND SO ON THIRD TO LAST LAST CHOICE LAST CHOICE SECOND TO LAST AND SO ON AND SO ON LAST CHOICE TESTING MENU TESTING MENU TESTING M...

Страница 18: ...e smaller end of the USB cable USB mini B connector into the meter and the larger end of the USB cable USB Type A connector into a USB port on a computer The battery icon will show no bars and ash whe...

Страница 19: ...Y REPLACEMENT The lithium ion battery used in this unit should last for many years with normal use When it no longer powers the unit long enough to meet testing requirements it will need to be replace...

Страница 20: ...tter with the return authorization number meter serial number a brief description of problem and contact information including phone and FAX numbers to the shipping carton This information will enable...

Страница 21: ...keypad These allow the user to select options from the menu driven software to directly read test results or to review stored results of previous tests in the data logger The menus can be displayed i...

Страница 22: ...e bottom The header displays the title of the current menu The footer line displays the time and the date the data logger status and the battery status The menu selection window is in the middle of th...

Страница 23: ...messages or the LaMotte logo is displayed The top two lines of the display are used to display information in a large easy to read format The menus work in the same way as previously described but tw...

Страница 24: ...use to remove dirt or ngerprints The tubes should be allowed to air dry in an inverted position to prevent dust from entering the tubes Dry tubes should be stored with the caps on to prevent contamina...

Страница 25: ...er tube and follow the test procedure Repeat the dilution and testing procedures until the result falls within the range of the calibration Multiply the test result by the dilution factor For example...

Страница 26: ...tomatic wavelength selection Languages English Spanish French Portuguese Italian Chinese Japanese Temperature Operation 0 50 C Storage 40 60 C Operation Humidity Range 0 90 RH non condensing USB Port...

Страница 27: ...ank Scan Sample 12 00 00 001 500 CALIBRATION As with all pre calibrated meters it is highly recommended even if not required by regulations that the user periodically verify the performance of the met...

Страница 28: ...er Shut Down Low Batt Shut Down Battery is too low to operate the unit Charge battery or use USB wall computer adapter Overrange Sample is outside of acceptable range Dilute sample and test again Unus...

Страница 29: ...3242 MN 7 19 www com information itm com 1 800 561 8187...

Отзывы: