4
4
DPD 3 - Code 6197A-J
*Chlorine Reagent #2 - Code 4499WT-H
ULTRA HIGH RANGE TOTAL CHLORINE
ColorQ Range: 0 -750 ppm
1. Fill sample bottle with water sample. Replace cap.
2. Fill tube to the 5 mL line with water sample.
8. Invert 1 time to mix. Insert tube into ColorQ.
6. Add 1
DPD 3
Tablet. Cap. Shake for TWENTY seconds. Invert slowly
5 times.
7. Allow tube to sit for TWO minutes. Tap tube on surface 3 times to remove
air bubbles.
9. Press button to read “uCL” Ultra High Range Total Chlorine.
10. Press button to return to blank meter or hold button until OFF is displayed.
Remove tube.
4. Press button to turn meter on. When “bLA” appears press button to
“Blank” the meter, press button FOUR times to go to “uCL” Ultra High
Range Total Chlorine. Remove tube.
3. Insert tube into ColorQ with arrow facing forward.
5. Add 5 drops
Chlorine #2
to same tube.
* Reagent is a potential health hazard. Read SDS at www.lamotte.com.
1
Color
Q
2
UCL
UCL