to return to bLA (blank) or hold button until OFF is displayed. Remove tube.
1. Add 1
DPD 1
Tablet to same tube.
3. Press button to read “FCL” Free Chlorine.
2. Cap. Shake for TWENTY seconds. Invert slowly 5 times. Tap tube on flat
surface 3 times to remove air bubbles. Insert tube into ColorQ.
4. Press button to go to “tCL” Total Chlorine. Remove tube.
Or press button past “tCL” (Total Chlorine) and “
UCL
” (Utlra High Range Chlorine)
3
* Reagent is a potential health hazard. Read SDS at www.lamotte.com.
*Hardness 1 Buffer: Code 7045-G
*Hardness 2 Indicator: Code 7046-G
3
b
DPD 3 - Code 6197A-J
cl
ColorQ Range: 0-10 ppm Chlorine
TOTAL CHLORINE
1
c l
Color
Q
2. Add 1
DPD 3
Tablet to the same tube. Cap. Shake for TWENTY seconds.
Invert slowly 5 times. Tap tube on flat surface 3 times to remove air bubbles.
5. Press button past “
UCL
” (Ultra High Range Chlorine) to return to bLA (blank)
or hold button until OFF is displayed. Remove tube.
3. Insert tube into ColorQ.
1. Remove cap from reacted FCL tube.
4. Press button to read “tCL” Total Chlorine.
FREE CHLORINE
pH
pH - Code 7059-E
ColorQ Range: 0-10 ppm Chlorine
2
fcl
DPD 1 - Code 6903A-H
F C L
Color
Q
1