28

 

„

SAMPLE DILUTION TECHNIQUES

If a test result is out of the range of the meter, it must be diluted. The test 

should then be repeated on the diluted sample. The following table gives quick 

reference guidelines for dilutions of various proportions.

Amount of Sample

Deionized Water to Bring 

Final Volume to 10 mL

Multiplication Factor

10 mL

0 mL

1

5 mL

5 mL

2

2.5 mL

7.5 mL

4

1 mL

9 mL

10

0.5 mL

9.5 mL

20

All dilutions are based on a final volume of 10 mL, so several dilutions will 

require small volumes of the water sample. Graduated pipets should be used for 

all dilutions. If volumetric glassware is not available, dilutions can be made with 

the colorimeter tube. Fill the tube to the 10 mL line with the sample and then 

transfer it to another container. Add 10 mL volumes of deionized water to the 

container and mix. Transfer 10 mL of the diluted sample to the colorimeter tube 

and follow the test procedure. Repeat the dilution and testing procedures until 

the result falls within the range of the calibration. Multiply the test result by the 

dilution factor. For example, if 10 mL of the sample water is diluted with three 

10 mL volumes of deionized water, the dilution factor is four. The test result of 

the diluted sample should be multiplied by four.

GENERAL INFORMATION

 

„

PACKAGING AND DELIVERY

Experienced packaging personnel at LaMotte Company assure adequate 

protection against normal hazards encountered in transportation of shipments.
After the product leaves LaMotte Company, all responsibility for safe delivery 

is assured by the transportation company. Damage claims must be filed 

immediately with the transportation company to receive compensation for 

damaged goods.

 

„

GENERAL PRECAUTIONS

READ THE INSTRUCTION MANUAL BEFORE ATTEMPTING TO SET UP OR OPERATE 

THE METER.

 Failure to do so could result in personal injury or damage to 

the meter. The meter should not be used or stored in a wet or corrosive 

environment. Care should be taken to prevent water from wet tubes from 

entering the meter chamber.

NEVER PUT WET TUBES IN THE METER.

Содержание 3684-01

Страница 1: ...DC1500 UREA Urea Colorimeter Kit Code 3684 01...

Страница 2: ...2...

Страница 3: ...ck 15 Logging Data 16 Factory Setup 17 Setting Power Save 17 Setting Backlight Time 18 Selecting A Language 19 Looping Menus 20 Computer Connection Output 21 Computer Connection 21 Battery Battery AC...

Страница 4: ...ing and Delivery 27 General Precautions 27 Safety Precautions 28 Limits of Liability 28 Specifications 29 CE 30 IP67 30 Warranty 30 Trouble Shooting Guide Error Messages 31 Calibration 31 Stray Light...

Страница 5: ...WARNING Reagents marked with an are considered to be potential health hazards To view or print a Safety Data Sheet SDS for these reagents go to www lamotte com Search the four digit reagent code numb...

Страница 6: ...trient for algae and bacteria It is introduced by bather s sweat and is the main component of urine APPLICATION Swimming Pools RANGE 0 0 6 0 mg L urea METHOD Urease catalyzes the hydrolysis of urea to...

Страница 7: ...on Main Menu Testing Menu Editing Menu 12 00 00 001 500 3 Press ENTER to select Testing Menu All Tests 096 Urea 12 00 00 001 500 4 Scroll to and select 096 Urea from menu 096 Urea Scan Blank Scan Sam...

Страница 8: ...strip up and down for 30 seconds Avoid hitting the bottom of the tube Remove and discard strip 8 Wait exactly 5 minutes 05 00 9 At the end of 5 minute waiting period use the 1 0 mL plastic pipet 0354...

Страница 9: ...dissolves Do not shake vigorously 13 Wait exactly one minute 01 00 14 At the end of one minute waiting period use the 0 1 g spoon 0699 to add two level measures of Salicylate 3 Reagent 7458 To add a l...

Страница 10: ...the test is performed To use the previous blank reading instead of scanning a new one scroll to Scan Sample and proceed For the most accurate results the meter should be blanked before each test and...

Страница 11: ...ION PROCEDURE 1 Press and briefly hold to turn the meter on The LaMotte logo screen will appear for about 3 seconds and the Main Menu will appear Main Menu Testing Menu Editing Menu 12 00 00 001 500 2...

Страница 12: ...strip up and down for 30 seconds Avoid hitting the bottom of the tube Remove and discard strip 7 Wait exactly 5 minutes 05 00 8 At the end of 5 minute waiting period use the 1 0 mL plastic pipet 0354...

Страница 13: ...or 30 seconds or until powder dissolves Do not shake vigorously 12 Wait exactly one minute 01 00 13 At the end of one minute waiting period use the 0 1 g spoon 0699 to add two level measures of Salicy...

Страница 14: ...Calibrate 096 Urea 0 99 mg L T Abs Calibrate 12 00 00 001 500 18 Press ENTER to select Calibrate A reverse font light background with dark characters will appear to indicate that the reading can be a...

Страница 15: ...ithout saving the adjustment press EXIT Press and ENTER at any time to return to the default value The calibration has now been standardized and can be used for testing Scroll to Scan Blank and begin...

Страница 16: ...500 2 Press ENTER to select Set Clock The year is displayed Press or to scroll to the appropriate character Press ENTER to select the character The month day hour format hour minute second AM PM will...

Страница 17: ...iting Menu press or to scroll to Logging Editing Menu Set Clock Logging Factory Setup Set Power Save 12 00 00 001 500 2 Press ENTER to select Logging Logging Display Test Log Logging Enabled Logging D...

Страница 18: ...he Editing Menu press or to scroll to Set Power Save Editing Menu Set Clock Logging Factory Setup Set Power Save 12 00 00 001 500 2 Press ENTER to select Set Power Save Set Power Save Disable 5 Minute...

Страница 19: ...the backlight is on the more frequently the battery will have to be charged if the USB Wall Adapter is not being used 1 From the Editing Menu press or to scroll to Backlight Time Editing Menu Logging...

Страница 20: ...001 500 3 Press or to scroll to desired language Select Language English Spanish French Portugese 12 00 00 001 500 4 Press ENTER The screen will display Storing for about 1 second and the meter will r...

Страница 21: ...rolling in either direction The diagrams below demonstrate a looping menu AND SO ON AND SO ON AND SO ON THIRD TO LAST LAST CHOICE LAST CHOICE SECOND TO LAST AND SO ON AND SO ON LAST CHOICE TESTING MEN...

Страница 22: ...B or actively charging by USB could result in damage to internal components To charge the battery from a computer plug the smaller end of the USB cable USB mini B connector into the meter and the larg...

Страница 23: ...utton Control and keep the backlight off BATTERY REPLACEMENT The lithium ion battery used in this unit should last for many years with normal use When it no longer powers the unit long enough to meet...

Страница 24: ...d recycling systems contact local or regional waste administration or recycling services GENERAL OPERATING INFORMATION OVERVIEW The meter is a portable microprocessor controlled direct reading colorim...

Страница 25: ...or selections The display has a header line at the top and a footer line at the bottom The header displays the title of the current menu The footer line displays the time and the date the data logger...

Страница 26: ...information such as test results or error messages or the LaMotte logo is displayed The top two lines of the display are used to display information in a large easy to read format The menus work in th...

Страница 27: ...nts The tubes should be allowed to air dry in an inverted position to prevent dust from entering the tubes Dry tubes should be stored with the caps on to prevent contamination After a tube has been fi...

Страница 28: ...e dilution and testing procedures until the result falls within the range of the calibration Multiply the test result by the dilution factor For example if 10 mL of the sample water is diluted with th...

Страница 29: ...Puerto Rico locations outside the North American Continent 813 248 0585 call collect Be prepared to supply the name and four digit LaMotte code number found on the container label or at the top of the...

Страница 30: ...Tests YES with automatic wavelength selection Languages English Spanish French Portuguese Italian Chinese Japanese Turkish Temperature Operation 0 50 C Storage 40 60 C Operation Humidity Range 0 90 RH...

Страница 31: ...800 344 3100 or 1 410 778 3100 ext 3 or softwaresupport lamotte com for a return authorization number or visit www lamotte com for troubleshooting help The sender is responsible for shipping charges...

Страница 32: ...e displayed in the test log 096 Urea Overrange Enter continue Scan Blank Scan Sample 12 00 00 001 500 CALIBRATION As with all pre calibrated meters it is highly recommended even if not required by reg...

Страница 33: ...wall computer adapter Shut Down Low Batt Shut Down Battery is too low to operate the unit Charge battery or use USB wall computer adapter Overrange Sample is outside of acceptable range Dilute sample...

Страница 34: ...34...

Страница 35: ...35...

Страница 36: ...ton Ave Chestertown Maryland 21620 USA 410 778 3100 800 344 3100 www lamotte com LaMotte WaterLink Spin Touch are registered trademarks of LaMotte Company 2021 LaMotte Company All Rights Reserved 3684...

Отзывы: