19. Press
ENTER
to select
Set
Calibration and save the
calibration. Press
or
to scroll and select
Factory Setting to revert to
the factory calibration. The
meter will momentarily display
Storing... and return to the
Turbidity -Without Blank
menu. The calibration has
now been saved and the
meter can be used for testing.
Turbidity WB (F)
Scan Blank
Scan Sample
12:00:00 001/500
NOTE: For the greatest accuracy during the calibration procedure,
be sure that after the meter is blanked and the blank is scanned as a
sample, the reading is 0.00. If not, reblank the meter and scan the blank
again until it reads 0.00. When scanning the calibration standards as
the sample, scan the calibration standard three times removing the
tube from the chamber after each scan and reinserting the tube in the
chamber with the same orientation. The readings should be consistent.
Use the last consistent reading to calibrate the meter. If the readings are
not consistent, avoid using an aberrant reading to calibrate the meter.
ANALYSIS WITHOUT BLANKING PROCEDURE
To obtain the most accurate results the meter should be blanked before
measuring a sample. The blanking step is not as critical for samples
above 10 NTU. The meter should always be blanked before reading
samples below 10 NTU.
1. Press and briefl y hold
to turn the meter on. The
LaMotte logo screen will
appear for about 3 seconds
and the
Main Menu will
appear.
Main Menu
Measure
Data Logging
Options
Run PC Link
12:00:00 001/500
2. Press
ENTER
to select
Measure.
Measure Menu
Turbidity - No Blank
Turbidity - With Blank
12:00:00 001/500
41