3. Remove
the
fi lter from the
syringe. Remove the plunger
from the barrel. (This step is
required to prevent rupturing
the fi lter by the vacuum that
would be created when the
plunger is removed.)
4. Replace
the
fi lter and repeat
step 2 for a second rinse
of the syringe and storage
container.
5. Remove
the
fi lter from the
syringe. Remove the plunger
from the barrel. Replace
the fi lter and fi ll the syringe
with approximately 50 mL
of deionized water. Filter
the water into the storage
container and save this
turbidity-free water.
6. Repeat Step 5 until the
desired amount of turbidity-
free water has been collected.
TESTING TIPS
1. Samples should be collected in a clean glass or polyethylene
container.
2. Samples should be analyzed as soon as possible after collection.
3. Gently mix sample by inverting before taking a reading but avoid
introducing air bubbles.
4. For the most precise results, follow the recommended procedure
48
Analysis
Содержание 2020we
Страница 1: ...2020we wi 1970 EPA 1970 ISO...
Страница 2: ......
Страница 57: ...57...
Страница 58: ...58...
Страница 59: ...59...
Страница 60: ...802 Washington Ave Chestertown Maryland 21620 USA 410 778 3100 800 344 3100 www lamotte com 1970 MN 02 01 12...