background image

Basic Operation

Powering the TRACER

If the batteries are weak, the

BAT

indicator will appear on the LCD.

Press the ON/OFF key to turn the TRACER on or off. The auto power
off feature will shut the TRACER off automatically after 10 minutes of
inactivity.

Electrode Recognition

When the TRACER is turned on, it will recognize the type of electrode
that is connected and will display the appropriate unit of measure. An
electrode must be attached before turning the meter on.

Automatic Calibration

When the TRACER is turned on, it will enter the Automatic
Calibration mode. ‘SELF’ and ‘CAL’ will appear on the display while
calibration is in progress. After the calibration is completed, the ‘SELF’
and ‘CAL’ display icons will extinguish and both the main display and
the bargraph will read ORP in mV.

Data Hold

Press the MODE/HOLD button to freeze the current reading. The

HOLD

display icon will appear along with the held reading. The held reading
will be stored in the next 15-reading storage location. Press the
MODE/HOLD button to return to normal operation.

TESTING

Overview

ORP is the abbreviation for Oxidation Reduction Potential and
represents the oxidizing or reducing nature of a solution. The overall
reactive tendency of a solution can be determined from ORP
measurements.

Display

When the electrode is placed in a solution, the main display and bar
graph will indicate the ORP reading. Readings flash until they have
stabilized. The bar graph is ‘center zero’, i.e. at 500mV there is no
display. As the readings increase from 0 to 1000 mV the bar will move
from the left to the right.

Getting Started
1.

Remove the cap from the bottom of the TRACER to expose the
flat surface electrode and reference junction.

2.

White KCL crystals may be present in the cap. Rinse the crystals
with tap water.

6

Содержание 1742 PockeTester

Страница 1: ...TRACER POCKETESTER ORP CODE 1742...

Страница 2: ......

Страница 3: ...asic Operation Powering the TRACER 6 Electrode Recognition 6 Automatic Calibration 6 Data Hold 6 ORP Testing Overview 6 ORP Display 6 Getting Started 6 ORP Measurement 6 Storing Readings 7 Recalling S...

Страница 4: ...unction LCD with Bar graph Operating Conditions 32 to 122 F 0 to 50 C and 80 RH ORP Range Accuracy 999mV 4mV Measurement storage 15 tagged numbered readings Power Four CR2032 button batteries see Page...

Страница 5: ...Battery compartment cap 2 LCD Display 3 MODE HOLD button 4 CAL RECALL button 5 ON OFF button 6 Electrode Collar 7 Electrode Note The Electrode cap is not shown Display Description 1 Reading 2 Units 3...

Страница 6: ...Hold Press the MODE HOLD button to freeze the current reading The HOLD display icon will appear along with the held reading The held reading will be stored in the next 15 reading storage location Pre...

Страница 7: ...epeat step 1 to store the next reading 4 If an attempt is made to store more than 15 readings the stored readings starting with the first reading will be overwritten Recalling Stored Readings NOTE Fir...

Страница 8: ...tore vertically Battery Replacement 1 Twist off the battery compartment cap 2 Hold the battery housing in place with one finger Remove the battery carrier by pulling on the small tabs 3 Replace the fo...

Страница 9: ...r and stir bars Display frozen HOLD function Press MODE or turn meter off Button response Remove batteries stored data will be lost Steady 1 display Wait Reading not stable yet Replacing the Electrode...

Страница 10: ...of one 1 year from the original purchase date The probe is guaranteed to be free from defects in material and workmanship for a period of six 6 months from the original purchase date In the event that...

Страница 11: ...LaMotte Company will service and repair out of warranty products at a nominal charge...

Страница 12: ...aMOTTE COMPANY Helping People Solve Analytical Challenges PO Box 329 Chestertown Maryland 21620 USA 800 344 3100 410 778 3100 Outside U S A Fax 410 778 6394 Visit us on the web at www lamotte com 5 08...

Отзывы: