![Lambrecht sun[e] 00.16130.501030 Скачать руководство пользователя страница 6](http://html.mh-extra.com/html/lambrecht/sun-e-00-16130-501030/sun-e-00-16130-501030_manual_671567006.webp)
6
1 Ordering and checking at delivery
1.1 Ordering sun[e]
Id-No. for standard configurations sun[e]:
Version 00.16130.501030:
digital ISO 9060 secondary standard pyranometer sensor with Modbus over RS-485
Common options / accessories are:
•
cable 5m, M12 plug connector
Id-No. 32.14567.060030
•
cable 12m, M12 plug connector
Id-No. 32.14567.060000
•
cable 15m, M12 plug connector
Id-No. 32.14567.060010
•
cable 20m, M12 plug connector
Id-No. 32.14567.060040
Suitable data loggers:
•
met[LOG]
Id-No. 00.95800.010000
•
Ser[LOG]
Id-No. 00.95770.000000
1.2 Included items
Arriving at the customer, the delivery should include:
•
pyranometer sun[e]
•
sun screen
•
cable of the length as ordered
•
product certificate matching the instrument serial number, including:
o
calibration certificate
o
temperature response test report
o
directional response test report
o
tilt sensor test report
•
any other options as ordered
Please store the certificates in a safe place.
1.3 Quick instrument check
1. At power–ON the signal may have a temporary output level different from zero; an offset. Let
this offset settle down; it is a normal part of the power-ON procedure.
2. Check if the sensor reacts to light: expose the sensor to a strong light source, for instance a
100 W light bulb at 0.1 m distance. The signal should read > 100 W/m
2
now. Darken the
sensor either by putting something over it or switching off the light. The instrument irradiance
output should go down and within one minute approach 0 W/m
2
.
3. Inspect the instrument for any damage.
4. Check the instrument serial number as indicated by the label on the instrument against the
certificates provided with the instrument.