LOW POWER OPERATING MODE
Operating condition
heater [OFF] and ventilator [OFF]
Zero offset a
5 W/m
2
Supply voltage range
5 to 30 VDC
Power consumption
< 0.1 W
CALIBRATION
Calibration traceability
to WRR
Calibration hierarchy
from WRR through ISO 9846 and ISO 9847, applying a
correction to reference conditions
Calibration method
indoor calibration according to ISO 9847, Type IIc
Calibration uncertainty
< 1.2 % (k = 2)
Recommended recalibration interval
2 years
Reference conditions
20 °C, normal incidence solar radiation, horizontal mounting,
irradiance level 1000 W/m
2
, heater and ventilator [ON]
Validity of calibration
based on experience the instrument sensitivity will not change
during storage. During use under exposure to solar radiation
the instrument “non-stability” specification is applicable.
MEASUREMENT ACCURACY AND RESOLUTION
Uncertainty of the measurement
statements about the overall measurement uncertainty can
only be made on an individual basis.
WMO estimate on achievable accuracy for
daily sums
2 %
WMO estimate on achievable accuracy for
hourly sums
3 %
Irradiance resolution
0.05 W/m
2
DIGITAL COMMUNICATION
Digital output
irradiance in W/m²
House keeping data
instrument body temperature in °C
instrument internal humidity in %r h
instrument internal pressure in Pa
instrument tilt angle in °
ventilator speed in RPM
ventilator current in A
heater current in A
Rated operating voltage range
8 to 30 VDC
Communication protocol
Modbus over 2-wire RS-485
half duplex
Transmission mode
RTU