Lambrecht METEOROLOGY 14576-24 V Скачать руководство пользователя страница 4

4

(145x6) 

METEOROLOGY 

   Operating 

Instructions

Operating Instructions
Wind Sensors METEOROLOGY (145x6)

If the preferred current of the power supply is according to 
point 1 this will be a current range of 6.0...7.45 mA.

The data defi ned in DIN 19 234 show that the triggering of 
switching or monitoring processes depends on the current in 
the interface between the proximity switch and the amplifi er. 
Thus, a current evaluation should be made when designing an 
amplifi er for proximity switches according to DIN 19 234.

The cable resistance must not exceed 100 ohms.

Example: Discrete circuit

 

Circuit with OP amplifi er

The electronic heating...

... of the 14576 24V and 14566 24V models is powered by 
an external power supply unit. The necessary voltage of
24 V

DC

 is connected via the wires M(-) and G(+).

Grounding

To reduce the risk of inductive interference the sensor must 
be properly grounded (screening on both sides).

The use of ready-made connecting cables (order no. 
32.14566.060 xxx) guarantees a proper grounding between 
sensor and cable. You just have to connect the cable screen-
ing to the ground of the indicator or the data acquisition unit. 

When other cables are used, you must create an appropriate 
screening. 

Setting up the North direction for the wind vane

For wind direction measurements the north mark on the sen-
sor must be aligned with the geographical north direction; the 
allowed maximum alignment deviation is ±3 degrees.

To set up the sensor’s north orientation select a landmark 
which is as far as possible up north with regard to the fi nal 
position of the wind direction sensor.

The reference point can be selected using a topographical 
map (1:25000). The exact position of the reference point is 
determined using an amplitude compass that can be adjusted 
horizontally on a stand.

A point on the rotary knob and another point on the casing 
shaft represent the north mark of the wind direction sensors. 
These two marks must be exactly in line with each other; when 
you have aligned the marks, you may fi x the wind vane with 
e.g. a piece of adhesive tape. When you have fi xed the wind 
vane this way you can locate the reference point by aiming 
at it over the axis. Now you must turn the sensor casing on 
the mounting tube until the tip of the wind vane points to the 
reference point in the north.

Please make sure there is no magnetic deviation of the 
compass.

Maximum interval 

Wind direction  Wind speed

a) 1 week 

Time mark of the measuring station
(correction if necessary)

 

Visual control of the device function

 Rough 

verifi cation of the measured 

values

b) 6 months 

Measured value transmission test

 

Verify that the sensors are still fi xed to 
the device carrier

 

  

Wind speed com-

 parison 

measurement 

 

using a hand-held

 anemometer

c) 2 years 

Sensor calibration

 Verifi cation of the north setup

If the weekly check according to a) gives cause to assume 
the existence of any malfunction you must perform the checks 
listed under b).

If the malfunctions are confirmed by these checks or are 
revealed for the fi rst time you must perform the checks listed 
under c).

When you have performed the checks listed under c) and the 
data are not measured correctly, please contact the manufac-
turer Wilh. LAMBRECHT GmbH.

When the north direction is set up for the wind direction sen-
sor, tighten the headless screws in the lower part of the casing 
and remove any adhesive tape, if necessary.

Using an indicating instrument or a data acquisition system, 
verify the proper measurement and transmission of the meas-
ured values for at least three directions with an offset of 90 
degrees from each other.

If you cannot select a northern reference point owing to local 
conditions, you can proceed analogously using a reference 
point in the south. In this case, however, you have to make 
sure the north mark on the sensor does not point to the refer-
ence point but in the opposite direction.

Maintenance

The sensor design permits long periods of maintenance-free 
operation. To obtain useful measured data the measuring sta-
tion must be systematically controlled and monitored.

The properties of the sensor may change due to wear, soil-
ing, and corrosion and as a result of external mechanical 
infl uences. The properties of the electronic module may also 
change.

We therefore recommend to check the measuring station pe-
riodically according to intervals listed in the following table:

Содержание METEOROLOGY 14576-24 V

Страница 1: ...with extremely low friction to detect rotations this also guarantees a long service life Cup Anemometer The three armed cup anemometer with a diameter of 215 mm is rotated by the wind The number of re...

Страница 2: ...ny mutual influence of the sensors A wind sensor pair easily meets this requirement since the sensors are set up side by side Their horizontal distance should be approximately 1 5 m The two sensors mu...

Страница 3: ...nt dependent switching or monitoring points 2 a Switching point The amplifier switching point must be within a proximity switch current consumption range of 1 2 mA to 2 1 mA 2 b Cable damage monitorin...

Страница 4: ...the wind vane this way you can locate the reference point by aiming at it over the axis Now you must turn the sensor casing on the mounting tube until the tip of the wind vane points to the reference...

Страница 5: ...ignement Id No 32 14565 060 xxx Coupler socket Id No 65 53070 470 100 View of the socket from the solder side View of the socket from the plug side View of a cable end The single leads of the cabel ar...

Страница 6: ...6 145x6 METEOROLOGY Operating Instructions Operating Instructions Wind Sensors METEOROLOGY 145x6 14576 I Id No 00 14576 010 000 14576 24V Id No 00 14576 250 004 Dimensions...

Страница 7: ...14566 24V Id No 00 14566 200 30 Wind tunnel results DWD Hamburg Wind speed sensor 14576 24V Instantaneous value v Measured frequence f m s Hz 0 30 0 1 07 7 5 2 10 16 5 4 95 45 9 82 93 14 75 142 19 80...

Страница 8: ...cup rotor 210 mm R80 Weight approx 0 4 kg approx 0 4 kg Other delivered with 12 pin plug and integrated transducer delivered with 12 pin plug No 00 14566 200 304 00 14566 000 030 Code 14566 24V 14566...

Отзывы: