1. 

Initial operation

The wind can be represented by a vector quantity. For a com-
plete description of the wind it is necessary to specify its speed 
and direction. The two components are subject to spatial and 
temporal variations; thus, strictly speaking, they are valid only 
for the site where the measuring instrument is put up. We there-
fore recommend to select the place of installation very carefully.

Selecting the place of installation

Generally, wind measuring instruments should not measure 
the specifi c wind conditions of a limited area, but indicate the 
typical wind conditions of a wider area. The values measured 
at diff erent places must be comparable. Thus, when installing 
the sensor you should make sure the place of installation is 
not under the lee of great obstacles. The distance between the 
obstacles and the sensor should be 10 times the height of the 
obstacles (this corresponds to the defi nition of an undisturbed 
terrain). 
If an undisturbed terrain of this kind does not exist the sensor 
must be put up at a height of at least 5 m above the obstacle 
height.
If the sensor must be installed on a roof top the place of instal-
lation must be in the middle of the roof to avoid predominant 
wind directions. If you want to measure both wind direction 
and wind speed, install the sensors at the same measuring 
point, if possible, and make sure to avoid any mutual infl u-
ence of the sensors. A wind sensor pair easily meets this 
requirement since the sensors are set up side by side. Their 
horizontal distance should be approximately 1.5 m. The two 
sensors must be staggered vertically so that the lower edge of 
the upper wind speed sensor is 0.1 to 0.5 m above the upper 
edge of the lower wind direction sensor.

2. 

Principles of installation

Attention!
Because the installation takes place in a danger-
ous height, the assembly personal must follow 
the rules for prevention of accidents.

I. Traverse with bore (e.g. Id-No. 32.14627.010000)

Material thickness for installation of the sensor between the 
nuts may be max. 10 mm.
2.   The sensor is led without cable into the bore and fastened 

by the opposite side with the loose nut. 

3.  Attach the sensor with the fl at side of detached nut from 

the lower side. Tighten with a suitable tool (wrench size 
36), until a twisting safety of the sensor aligned to the north 
is given.

II. Mast or pipe mounting

Make sure the device is easily accessible so that you 
can set up the north direction for the wind direction 
sensor and perform any maintenance work. To reach 
the sensors use a ladder of the appropriate length 
or a telescoping working platform of the appropriate 
height.

Ladders or other lifting helps must be absolutely 
in order and must be guarantee a secure support! 
Follow the rules for prevention of accidents.

Mount the sensors at the top of grounded tube with an outer 
diameter Ø 48-50 mm. The mast adapter (see accessories) 
is obligatory.
1.  Remove both thread nut from the sensor. 
2.  The sensor is inserted without cable into the bore 

(Ø 30 mm) of the adapter and locked from the opposite 
side with a loose nut in the direction of the adapter. 

3.  And fi nally we recommend to lock the second nut with its 

plane side ahead against the fi rst nut (see drawing).

If wind speed and wind direction are measured at the same 
time, the measurement generally takes place not only at the 
top of a mast but also at the ends of a cross arm. The arms 
must stay torsion-free and vibration-proof even at high wind 
speeds and they must be accessible for you to perform mount-
ing and maintenance work.

2

Содержание INDUSTRY Modbus

Страница 1: ...good starting values large measuring and temperature application ranges easy mast mounting optimal heating concept Content 1 Initial operation 2 Principles of installation 3 Setting up the North Dire...

Страница 2: ...lower edge of the upper wind speed sensor is 0 1 to 0 5 m above the upper edge of the lower wind direction sensor 2 Principles of installation Attention Because the installation takes place in a dang...

Страница 3: ...measurements the north mark on the sen sor must be aligned with the geographical north direction You have to turn the marking exactly over the marking at the sensor shaft When you have aligned the mar...

Страница 4: ...14567 14577 INDUSTRY Modbus Manual Wind Sensors INDUSTRY Modbus Manual 14567 14577 INDUSTRY Modbus Manual 7 Wiring diagram 4...

Страница 5: ...Manual Wind Sensors INDUSTRY Modbus 14567 14577 INDUSTRY Modbus Manual 8 Dimensioned drawings Accessories optional 14567 U6 Mast adapter Id No 32 14567 006 000 5...

Страница 6: ...s 0x1234 is transmitted in the order 0x12 0x34 Example Big Endian 32 bit or 64 bit Register size value 32 bits 0x12345678 is transmitted in the order 0x12 0x34 0x56 0x78 9 2 Device address The address...

Страница 7: ...al place INT 30202 Wind direction average value since last retrieval 10 1 decimal place INT 30203 Wind direction maximum value 10 1 decimal place INT 30204 Wind direction minimum value 10 1 decimal pl...

Страница 8: ...perature If the autocon guration is started with the data logger met LOG it queries the available mapping registers on each COM interface in the device address range 1 25 For this purpose the number o...

Страница 9: ...Endian WORD 31592 Global radiation instantaneous values Low WORD uncompensated 0x04 Big Endian WORD 9 8 1 Mapping register INDUSTRY Modbus Register address Register value Unit Factor Description 4600...

Страница 10: ...0 7 50 m s Accuracy 2 0 5 m s at 0 7 5 m s and 2 FS at 5 02 50 m s Resolution 1 0 02 m s Starting value 0 7 m s 0 7 m s Interface RS485 RS485 Protocol Modbus RTU Modbus RTU Weight 0 35 kg 0 25 kg Mea...