OPERATING INSTRUCTIONS 
METEODIGIT IV (9164) 

 

 

 

 6  

5. INTRODUCTION 

METEODIGIT IV is a new hand-held measuring instru-
ment of LAMBRECHT. The intelligent connector offers 
decisive advantages when connecting sensors because 
all parameters are stored in on the connector itself; 
repeat programming is thus no longer necessary. 

5.1 Functions 

METEODIGIT IV has one electrically isolated measur-
ing input suitable for all LAMBRECHT sensors. For 
operation purposes the device incorporates a large 
LCD display and a keypad. 

5.1.1 Sensor programming 

The connector programs the measuring channels com-
pletely and automatically. Programming of the connec-
tor will be done during manufacture.  
Accordingly programmed are:  

- Measuring 

ranges 

Dimension (2-digit)  

Correction of measured values 

- Scaling 

All sensor data stored in the connector can be pro-
tected by means of a graduated locking function 
against undesired access. 

5.1.2 Measuring operations 

The measurement input can be connected with double 
sensors, individually scaled sensors and sensors with 
function channels. You can move forwards or back-
wards from one measuring channel to the next using 
the keypad. The data are shown constantly on the dis-
play. 

Measured values 

The measured value for the selected measuring point is 
shown continuously with autozero and, as and when 
necessary, with measured value correction.  

Measuring functions 

With some sensors, to achieve optimal measured value 
acquisition, certain special measuring functions are 
required. For certain sensors temperature compensa-
tion and atmospheric pressure compensation are avail-
able.  

Maximum and minimum values 

For each measuring operation the maximum value and 
minimum value are acquired and saved to memory. 
These values can then be displayed, output, or deleted 
from memory. 

Measured value memory 

Up to 99 measured values can be saved manually. 
These data can be shown successively on the display. 

Differential measurement 

It is possible, by setting the measured value to zero, to 
perform relative measuring operations with respect to a 
reference value. 
 
 
 

6. INITIAL COMMISSIONING 

1.

  Connect sensor to socket 

 (M0) 

1

 

 

2.  Ensure that the power supply is provided with 3 AA 

batteries 

3.  Switch ON by pressing key 

 

ON

 

 

4

  

4.  Select measuring channels by pressing key 

 

M

 

4

, read out measured values (

3

e)  

5.  Save the measured values by pressing key 

 

MEM

 

 

4

 

6.  Relative measuring with respect to a reference 

value or sensor adjustment by pressing key

CLR

 

 

4

, return to normal measured value by pressing 

and holding down key 

 

CLR

 

  

7.  Evaluating a measuring operation  

Call up maximum / minimum values by pressing 
keys  

 

MAX

 

 

/

 

 

MIN

 

 

4

To delete max. / min. values press and hold down 
key 

 

MAX

 

 or 

 

MIN 

  

7. POWER SUPPLY 

Power can be supplied to the measuring instrument 
with 3 AA alkaline batteries (included in delivery). 

7.1 Battery operating and supply volt-
age monitoring  

At a current consumption of approx. 16 mA the operat-
ing time will be approx. 150 hours. The current operat-
ing voltage is displayed each time the device is 
switched on; this gives you a basis for estimating the 
remaining operating time. When the remaining battery 
capacity drops to approx. 10 percent, the LOBAT arrow 

 will appear on the display. If the batteries are 

completely discharged the device will switch off. To 
replace old batteries first unscrew the battery compart-
ment cover 

5

on the rear of the device. 

7.2 Switching ON / OFF, reinitialization 

To switch the device ON briefly press and release the 
key 

 ON  OFF

r

 

4

 in the middle of the keypad; to switch 

the device OFF press and hold down the key 

 ON  OFF

r

 . 

After the device is switched off all saved values and 
settings are retained intact.  

If interference (e.g. electrostatic) or a malfunction (e.g. 
battery failure) causes the device to behave abnor-
mally, the device can be reinitialized.  

To activate 

RESET 

press and hold down the key 

  CLR 

r

 

when switching on. This will restore all settings – except 
the device designation – to the factory default status.  

 

8. CONNECTING THE SENSORS 

All LAMBRECHT sensors can the connected to input 
socket 

 (M0) at the measuring instrument.  

Содержание 9164

Страница 1: ...and cost effective solution e g for production facilities quality management air conditioning technology and ventilating applications on ships and in scientific laboratories Digital measuring techniq...

Страница 2: ...tes Battery voltage 3 6V Unlocked for adjust purposes Measured value corrected Relative measuring Operating keys ON OFF Switch ON to switch OFF press long M M Meas point selection MAX MIN Max min valu...

Страница 3: ...onitoring 6 7 2 Switching ON OFF reinitialization 6 8 CONNECTING THE SENSORS 6 8 1 1 Sensors 7 8 1 2 Measuring inputs and additional channels 7 9 DISPLAY AND KEYPAD 7 9 1 1 Display 7 9 1 2 Keypad 7 10...

Страница 4: ...z The device is damaged by electrostatic dis charge or lightning z Improper handling respectively non observance of the operating instructions The manufacturer reserves the right to change the product...

Страница 5: ...ot use the device inside detonation areas or filling stations 4 1 1 Special notes on use If the device is brought into the work room from a cold environment there is a risk that conden sation might fo...

Страница 6: ...ement It is possible by setting the measured value to zero to perform relative measuring operations with respect to a reference value 6 INITIAL COMMISSIONING 1 Connect sensor to socket M0 1 2 Ensure t...

Страница 7: ...igits b for depicting various measuring func tions a there are also 4 arrows f for depicting the operating status Function field Main field Display of measuring functions in the function field Measuri...

Страница 8: ...d 55 Difference millivolt 1 mV d 26 Difference millivolt 2 mV d260 Difference volt V d2 60 Sensor voltage V UbAt Frequency Hz FrEq Function channels Maximum value of channel Mb1 Hi Minimum value of ch...

Страница 9: ...CLR Select the next digit or entry is completed ON To terminate or cancel configuration MEM 11 1 Moving average METEODIGIT IV offers all the possibilities to calculate the average value In the first...

Страница 10: ...e listed checks and reme dial steps the device still fails to behave as described in the operating instructions it must be returned to our factory in Goettingen with error description 13 DECLARATION O...

Страница 11: ...ly not separated Housing L 127 x B 83 x H 42 mm ABS weight approx 260 g Suitable conditions Operating temperature 10 50 C storage temperature 20 60 C Ambient relative humidity 10 90 r h non condensati...

Отзывы: