Wind Sensors EFFICIENT (14579)

1. 

Initial operation

The wind can be represented by a vector quantity. For a com-

plete description of the wind it is necessary to specify its speed 

and direction. The two components are subject to spatial and 

temporal variations; thus, strictly speaking, they are valid only 

for the site where the measuring instrument is put up. We there-

fore recommend to select the place of installation very carefully.

Selecting the place of installation

Generally, wind measuring instruments should not measure 

the specific wind conditions of a limited area, but indicate the 

typical wind conditions of a wider area. The values measured 

at different places must be comparable. Thus, when installing 

the sensor you should make sure the place of installation is 

not under the lee of great obstacles. The distance between the 

obstacles and the sensor should be 10 times the height of the 

obstacles (this corresponds to the definition of an undisturbed 

terrain). 
If an undisturbed terrain of this kind does not exist the sensor 

must be put up at a height of at least 5 m above the obstacle 

height.
If the sensor must be installed on a roof top the place of instal-

lation must be in the middle of the roof to avoid predominant 

wind directions. If you want to measure both wind direction 

and wind speed, install the sensors at the same measuring 

point, if possible, and make sure to avoid any mutual influ

-

ence of the sensors. A wind sensor pair easily meets this 

requirement since the sensors are set up side by side. Their 

horizontal distance should be approximately 1.5 m. The two 

sensors must be staggered vertically so that the lower edge of 

the upper wind speed sensor is 0.1 to 0.5 m above the upper 

edge of the lower wind direction sensor.

2. 

Principles of installation

Attention!
Because the installation takes place in a danger

-

ous height, the assembly personal must follow 

the rules for prevention of accidents.

I. Traverse with bore (e.g. Id-No. 32.14627.010000)

Material thickness for installation of the sensor between the 

nuts may be max. 10 mm.
2.   The sensor is led without cable into the bore and fastened 

by the opposite side with the loose nut. 

3.  Attach the sensor with the flat side of detached nut from 

the lower side. Tighten with a suitable tool (wrench size 

36), until a twisting safety of the sensor aligned to the north 

is given.

II. Mast or pipe mounting

Make sure the device is easily accessible so that you 

can set up the north direction for the wind direction 

sensor and perform any maintenance work. To reach 

the sensors use a ladder of the appropriate length 

or a telescoping working platform of the appropriate 

height.

Ladders or other lifting helps must be absolutely 

in order and must be guarantee a secure support! 

Follow the rules for prevention of accidents.

Mount the sensors at the top of grounded tube with an outer 

diameter Ø 48-50 mm. The mast adapter (see accessories) 

is obligatory.
1.  Remove both thread nut from the sensor. At sensor type 

with assembled cable to take off the nut over completely 

cable length. 

2.  The sensor with cable-plug connection is led without cable 

into the bore and fastened by the opposite side with a 

loose nut.  

3.  The sensors with assembled cable you have to pull 

through the cable beginning with the open cable end 

completely, until the sensor is on the edge of bore. 

4.  The loose nut to be touched now with the flat side first over 

the cable, in order to fasten the sensor. 

And finally we recommend to lock the second nut with its plane 

side ahead against the first nut (see drawing).

2