126
c
HAPTER
6:
Computer Interface Operation
Model 335 Temperature Controller
WARMUP
Warmup Supply Parameter Command
Input
WARMUP <output>,<control>,<percentage>[term]
Format
n,n,+nnn.nn
<output>
Output 2 is the only valid entry and must be included.
<control>
Specifies the type of control used: 0 = Auto Off,
1 = Continuous
<percentage>
Specifies the percentage of full scale (10 V) Monitor Out
voltage to apply to turn on the external power supply.
Example
WARMUP 1,50[term]
Output 2 in Voltage mode will use continuous control, with a
5 V (50.50%) output voltage for activating the external power supply.
Remarks
Warmup mode applies only to Output 2 in Voltage mode. The Output Type parameter
must be configured using the HTRSET command, and the Output mode and Control
Input parameters must be configured using the OUTMODE command.
WARMUP?
Warmup Supply Parameter Query
Input
WARMUP? <output>[term]
Returned
<control>,<percentage>[term]
Format
n,+nnn.nn (refer to command for description)
ZONE
Control Loop Zone Table Parameter Command
Input
ZONE <output>,<zone>,<upper bound>,<P value>,<I value>,
<D value>,<mout value>,<range>,<input>,<rate>[term]
Format
n,nn,+nnnnn, +nnnnn,+nnnnn,+nnnn,+nnnnn, n,n, +nnnn [term]
<output>
Specifies which heater output to configure: 1 or 2.
<zone>
Specifies which zone in the table to configure.
Valid entries are: 1–10.
<upper bound>
Specifies the upper Setpoint boundary of this zone in kelvin.
<P value>
Specifies the P for this zone: 0.1 to 1000.
<I value>
Specifies the I for this zone: 0.1 to 1000.
<D value>
Specifies the D for this zone: 0 to 200%.
<mout value>
Specifies the manual output for this zone: 0 to 100%.
<range>
Specifies the heater range for this zone. Valid entries:
0 = Off, 1 = Low, 2 = Med, 3 = High.
<input>
Specifies the sensor input to use for this zone. 0 = Default
(Use previously assigned sensor), 1 = Input A, 2 = Input B,
<rate>
Specifies the ramp rate for this zone: 0.1 –100 K/min.
Remarks
Configures the output zone parameters. Refer to Paragraph 2.9.
Example
ZONE 1,1,25.0,10,20,0,0,2,2,10[term]
Output 1 zone 1 is valid to 25.0 K with
P = 10, I = 20, D = 0, a heater range of medium, sensor input B, and a
ramp rate of 10 K/min.
ZONE?
Output Zone Table Parameter Query
Input
ZONE? <output>,<zone>[term]
Format
n,nn
<output>
Specifies which heater output to query: 1 or 2.
<zone>
Specifies which zone in the table to query.
Valid entries: 1–10.
Returned
< upper boundary>,<P value>,<I value>,<D value>,<mout
value>,<range>,<input>,<rate>[term]
Format
+nnnnn,+nnnnn,+nnnnn,+nnnn, +nnnnn,n,n, +nnnn
(refer to command for description)