CalCurve (formerly called Precision Option) is the easiest way to combine
the performance of a Lake Shore calibrated sensor with the Model 234. If
CalCurve is present, Place the
T/R
switch in the
T
position. See Paragraph
3.1 for output current temperature ranges.
The factory stores the optional Model 8001-234 CalCurve in the unit if it is
ordered with the unit. The Curve EPROM (see Figure 5-4) is labeled with the
sensor serial number. See Chapter 5 for Curve EPROM field installation.
The optional Model 8001-234 CalCurve data is stored in a temperature
versus log of resistance breakpoint table. The log of resistance is uniformly
spaced over three regions: log(R) of 0 to 2.98 is spaced every 0.02; log(R) of
3.0 to 3.96 is spaced every 0.04; and log(R) of 4.0 to 5.6 is spaced every
0.1. Six temperature digits are stored to the 1 mK place. A summary of curve
breakpoints, sensor resistance, log of resistance, and maximum temperature
error comes with each CalCurve. See Chapter 4 for Curve programming or
verifying curve data via the serial interface.
2.5
POWER CONNECTIONS
A +5 VDC supply in the VME rack or an external power supply powers the
Model 234. Regulate the voltage to within ±0.25 VDC. Each Model 234
draws up to 500 mA from the supply (750 mA for the Model 234D). The
external power supply connector must be S-760 or S-765 Switchcraft (or
equivalent) plug (0.218 inch O.D., accepts 0.08 inch diameter pin) with
+5 VDC on the sleeve and return on the center pin.
A wall plug-in power supply Model 2007-XX +5 VDC Regulated Power
Supply can be used with the Model 234. Power Supply input is based on
local power requirements as follows:
MODEL
INPUT POWER
2007-12
2007-22
120 V 60 Hz. power source
230 V 50 Hz. power source
CAUTION: Never ground both the sensor and the 4–20 mA output.
Ground either the sensor, or the output, but not both.
The Model 2308-10 VMEbus rackmount case has a built-in power supply for
use with up to twelve Model 234s. The built-in power supply has a universal
input: 85 to 265 VAC, 47 to 440 Hz, 60 watts.
Содержание 234
Страница 13: ...Lake Shore Model 234 234D User s Manual Introduction 1 9 Figure 1 3 Model 2308 1 Enclosure Physical Dimensions ...
Страница 14: ...Lake Shore Model 234 234D User s Manual 1 10 Introduction Figure 1 4 Model 2308 12 Case Physical Dimensions ...
Страница 20: ...Lake Shore Model 234 234D User s Manual 2 6 Installation Figure 2 2 Typical Wall Plug In Power Supply ...
Страница 24: ...Lake Shore Model 234 234D User s Manual 3 4 Operation This Page Intentionally Left Blank ...
Страница 36: ...Lake Shore Model 234 234D User s Manual 5 4 Service Figure 5 3 Model 234 PCB Details ...
Страница 44: ...Lake Shore Model 234 234D User s Manual 5 12 Service This Page Intentionally Left Blank ...
Страница 48: ...Lake Shore Model 234 234D User s Manual 6 4 Options Accessories This Page Intentionally Left Blank ...
Страница 49: ......