33
LOCK
Front Panel Keypad Lock Command
Input
LOCK<state>[term]
Format
n
<state>
0 = unlock, 1 = lock
Remarks
Allows for the user to lock or unlock the front panel
keypad
LOCK?
Front Panel Keypad Lock Query
Input
LOCK?[term]
Format
n (refer to command for description)
Returned
<state>{term]
PWUPENBL
Power Up Enable
Input
PWUPENBL<state>[term]
Format
<state>
0 = disabled, 1 = enabled
Remarks
Specifies whether the output remains on or shuts
off after power cycle.
RANGE
Current Range Command
Input
RANGE<range>[term]
Format
nn
<range>
0 = 100 nA
1 = 300 nA
2 = 1 µA
3 = 3 µA
4 = 10 µA
5 = 30 µA
6 = 100 µA
7 = 300 µA
8 = 1 mA
9 = 3 mA
10 = 10 mA
11 = 30 mA
12 = 100 mA
13 = User current
Remarks
Allows the user to change the current range. Use
the IPOL command to set the polarity.
Содержание 121
Страница 6: ......
Страница 10: ...iv ...
Страница 46: ...36 CHAPTER 5 Options and Accessories ...
Страница 54: ...44 CHAPTER 6 Service ...
Страница 60: ...50 APPENDICES B 1 1 CE Declaration for Model 121 ...
Страница 61: ...51 B 1 2 CE Declaration for the Power Supply Part Number 109 132 ...
Страница 62: ...52 APPENDICES ...