background image

No.:GC-JS-21                                    Version: 01/00                                       

Effective date

2020-07-22

 

 

 

Pressure Steam Sterilizer 

Autoclave LFSS03AA 

User Manual   

V1.0 

   

 
Manufacturer: 

Ningbo Haishu Life Medical Technology Co., Ltd. 
No. 1, Jinghui Road, Hengjie, Haishu, Ningbo 315181 China P.R.C. 
 

EC Representative: 

Caretechion GmbH 
Niederrheinstr 71,40474 Duesseldorf, Germany 

 

              Please appoint special person to operate and maintain the device. The 

operator and maintenance must be well trained 

                                                                                                                                           

Содержание Autoclave LFSS03AA

Страница 1: ...0 Manufacturer Ningbo Haishu Life Medical Technology Co Ltd No 1 Jinghui Road Hengjie Haishu Ningbo 315181 China P R C EC Representative Caretechion GmbH Niederrheinstr 71 40474 Duesseldorf Germany Pl...

Страница 2: ...No GC JS 21 Version 01 00 Effective date 2020 07 22 2 Revision History Revised chapter Revised content Revised date Revisor Revised number All New document 2020 7 22 Kejie Zheng V1 0...

Страница 3: ...stalltion 12 4 Adjusting 16 5 Introduction of preset programs 17 6 User interface description 19 Instructions of operation 25 8 Errors list 29 9 Maintenance 32 10 Quality assurance 37 11 Accessories 3...

Страница 4: ...manual must be transferred or handed over as part of the whole equipment Please read this manual carefully before using the equipment be familiar with the operation and safety instructions of the equi...

Страница 5: ...nteed Emergency handling If an emergency is found it shall be handled as follows Turn off the power switch Unplug Product responsibilities Without the written permission of the manufacturer it shall n...

Страница 6: ...r oils and powders such as petroleum jelly and agar This sterilizer cannot be used for sterilization of liquids or liquids sealed with closed containers especially glassware which may cause the contai...

Страница 7: ...he abnormal conditions chapter of the manual Necessary monitoring shall be carried out in accordance with relevant national and regional regulations After putting the monitoring substance such as chem...

Страница 8: ...tomatic protection ensure the effectiveness of sterilization and disinfection 1 1 Product categories The product is classified as class I type B equipment according to electrical safety The product is...

Страница 9: ...water vapor is between 134 and 137 and the temperature difference between each point in the sterilization chamber at the same time is no more than 2 and the maintenance time is 4min 2 sterilization f...

Страница 10: ...rameter of autoclave Max working temperature 137 Max working pressure 0 24MPa Safety valve set pressure 0 24MPa Safety valve open pressure 0 24MPa 0 26MPa Capacity of clean water tank 1L Waste water t...

Страница 11: ...heavy metals 10mg L 1mg L 0 2mg L 0 005mg L 0 05mg L 0 1mg L Chloride Phosphate Conductivity PH value Conductivity hardness 2mg L 0 5mg L 15uS cm 5 7 5 Colorless clean without precipitation 0 02mmol L...

Страница 12: ...hether it meets the requirements according to the following clauses 3 2 1 Power requirements AC220V 50Hz unidirectional power supply voltage fluctuation does not exceed 10 power 3 5kVA 3 2 2 Environme...

Страница 13: ...ich the sterilizer is placed is strong enough do not block the vent of the sterilizer 3 3 Handling of packaged products To ensure safe handling please use a cart and similar tools to move the device I...

Страница 14: ...t of the product it is placed on the workbench Please note the following when handling It is strictly forbidden to lift the door during transportation to avoid damage to the device Do not lift the foo...

Страница 15: ...rain Connect the pipe from the outlet of waste water chapter 1 4 to the sewers or storage containers 3 8 Insert the power plug into the socket Make sure that the equipment is reliably grounded 3 9 Oth...

Страница 16: ...k test The leak test is done to ensure the tightness of the sterilization chamber and its pipelines to ensure that the cold air in the sterilization chamber is eliminated This device comes with a leak...

Страница 17: ...aper plastic packaging or double layer packaging may not dry should be used immediately after sterilization Universal Non wrapped load or wrapped solid fabric and cavity load 134 solid Solid load with...

Страница 18: ...No GC JS 21 Version 01 00 Effective date 2020 07 22 18 min tion mL min tion mL min tion mL min tion mL LFSS0 3AA 2 2kg 20 280 24 300 38 380 18 240...

Страница 19: ...water quality 29 means that the current water quality is 29uS cm Shows whether the pre heating function is on On is on Off is not on Count 00001 indicates that it has been used once Display the curre...

Страница 20: ...ation and other information please refer to Chapter 5 of this manual 6 1 3 Settings You can select the setting menu on the main page and press OK to enter The program menu after entering is as The pro...

Страница 21: ...can press Start Stop to move to Month and so on after setting all dates and times press OK to save and exit 6 1 3 3 to switch different temperature units When cursor move to this option press OK butt...

Страница 22: ...djust to higher drying class to solve this issue When cursor move to this option press OK button to adjust the drying class I II III The drying time corresponding to each level is as follows 134 Unive...

Страница 23: ...he program once and water needs to be refilled 2 Need to close the door before running the program 3 Door is closed 4 The door is locked to ensure that the door cannot be opened during operation 5 Bad...

Страница 24: ...representative is required to maintain the equipment 10 Last cycel ended improperly Since the last sterilization cycle was stopped by the user the power failure or failure did not end normally the loa...

Страница 25: ...ing The equipment cannot run the program when the clean water tank is short of water and the water needs to be added to the water clean tank When there is no water in the equipment storage tank or whe...

Страница 26: ...ponding to each program see chapter 5 2 1 There is no difference in sterilization effect between the 134 C sterilization procedure and the 121 C sterilization procedure It is recommended to to use the...

Страница 27: ...you can turn the door handle counterclockwise to open the door 7 9 2 Offload After opening the door after sterilization the load can be taken out The operator should use the configured instrument tray...

Страница 28: ...erilization chamber is completely released 1 Use a tool to open the safety valve to completely release the internal pressure Be careful not to get burned by the discharged water vapor Can be operated...

Страница 29: ...32 Heating ring temperature 220 heating jacket temperature sensor damaged Check the temperature sensor in heating jacket 3 E33 Steam generator heating tube temperature 230 The temperature sensor of st...

Страница 30: ...ed Clean and sterilize indoor filters Check the drain solenoid valve vacuum selection valve 14 E6 The door switch is disconnected when the program is running The door switch sensor is displaced the mi...

Страница 31: ...eal 1 check if same result with empty loading 2 Check the vacuum pump 3 Check whether the door is closed enough to adjust the position of the door 21 E16 There is still a large negative pressure in th...

Страница 32: ...cessories 9 2 Maintenance plan 9 2 1 Cleaning plan No Cleaning items Cycle Requirement Remark 1 Sterilization chamber cleaning once a week Keep clean without sewage Refer to 9 3 1 2 Clean water tank O...

Страница 33: ...ine cleaning agents to clean the sterilization chamber Before cleaning the sterilization chamber please confirm that the sterilization room has been cooled to avoid burns 9 3 2 Cleaning the water stor...

Страница 34: ...sical program 2 Wait until the pressure is near 100kpa pull the pull ring on the safety valve to make it open wait for about 1s there will be air and steam exhausted indicating that the safety valve c...

Страница 35: ...it fails please contact the supplier or supplier representative The test results may be inaccurate when the sterilization room is hot or not dry and must be tested when the sterilization room is cold...

Страница 36: ...eft and right as shown in the figure below After softening press the remaining part into the sealing groove at a time and then insert the cover plate and tighten the four screws When replacing the sea...

Страница 37: ...itions are not covered by the free warranty even during the warranty period 1 Failure or damage caused by improper installation and use 2 Failure or damage caused by accidental drop or collision 3 Fai...

Страница 38: ...terilized items Place on the instrument rack 2 Instrument rack optional Sterilized items in the lower layer Upper instrument tray 3 External waste water tank optional To store the waste water generate...

Страница 39: ...Pass Empty load Inspection of finished product Pass Solid load Type test Pass Small amount of porous permeable strips Type test Pass Small amount of porous permeability load Type test Pass Fully poro...

Страница 40: ...02 RAISE 15 53 00 021 089 2 00 15 VACUUM 15 53 50 080 095 2 00 50 RAISE 15 54 05 010 103 9 00 14 VACUUM 15 55 09 080 105 7 01 05 RAISE 15 56 24 221 133 7 01 15 STER START 15 56 24 221 133 7 01 15 MAX...

Страница 41: ...injected into the sterilization chamber temperature and pressure will increase 5 drain The sterilization room is draining 6 Sterilization Sterilizing maintain sterilization temperature and pressure 7...