Lafayette 54035A Скачать руководство пользователя страница 8

Multifunction Timer / Counter

3700 Sagamore Parkway North  .  Lafayette, IN 47904 USA  .  Phone: (765) 423-1505

8

Timer Control Software Model 54035A-SF

The Multi-Function Timer/Counter is available with software that can be used to control the timer and collect 

and store data. The software interfaces to the timer via a standard serial connection. The software is built to 

send and receive data from the timer; no timing functions can be performed by the software itself. The software 

features a display that will show all time and count data gathered by the 54035A. It also has control buttons 

that mirror the buttons on the timer itself. In general, the software is designed to function exactly as the front 

panel control of the 54035A with the exception of it’s data storage capabilities.

Installation

** Windows 95 or higher is required to run the 54035A software.

1.  Insert Installation disk 1 into the A: floppy disk drive.
2.  Go to “Start”>>Run and type A:\setup.exe
3.  Insert disk 2 into the A: disk drive when prompted.
4.  Select a destination directory and click the computer icon to proceed with the installation.
5.  Insert disk 3 into the A: disk drive when prompted.
6.  To start the program, run 54035a.exe.

Software Functions

Comm port selection

On startup, the software will ask the user to designate which comm port to use with the 54035A. The selection 

will default to comm 2 as most computers use a mouse on comm 1. Selecting the comm port in use by the 

mouse will result in an error and/or disabling of the mouse. If the mouse is disabled, the computer will have 

to be restarted to restore function.

Timing functions

1. 

Start:

  Starts the timer.

2. 

Stop1: 

 Stops the timer, records split time 1 and updates the screen.

3. 

Stop2:

 

 Stops the timer, records split time 2 and updates the screen.

4. 

Stop3:

  Stops the timer, records split time 3 and updates the screen.

5. 

Stop4: 

 Stops the timer, records split time 4 and updates the screen.

6. 

Reset: 

 Clears the display, and sets all time and count data to zero. Reset can not be executed while the 

timer is timing.

7. 

Mode: 

 Enters the Mode menu on the timer and in the software. If the mode menu in entered from the 

software, all settings should be set changed from the software. If settings are entered manually on the 

timer while the mode menu is active on the software, errors in the settings may occur. The mode menu 

can not be entered unless the timer has been reset.

8. 

Update Data: 

 Transfers all time and count data from the 54035A to the computer screen. This is used when 

the times are collected with the timer controls and the software is only used to display or store the data.

9. 

Select File: 

 Allows the user to designate a file to store data gathered from the 54035A. All files are opened 

as “Append”, so files may be opened multiple times without data being overwritten. Data from new trials is 

added to the preexisting data. After the file is selected, the user will be prompted to enter a subject name 

or trial description. This entry will be used to identify results in the storage file.

Содержание 54035A

Страница 1: ...t com Model 54035A Multifunction Timer Counter User s Manual 3700 Sagamore Parkway North Lafayette IN 47904 USA Tel 765 423 1505 800 428 7545 Fax 765 423 4111 info lafayetteinstrument com www lafayett...

Страница 2: ...ion Timer Counter Model 54035A is designed to time and count events with high accuracy and high resolution It features millisecond timing up to 1000 seconds and counting up to 1000 counts It can take...

Страница 3: ...mode button on the front panel or via instructions on the serial communication port seeCommunications directions pg 9 for details In order to enter the mode menu the timer must be reset Single Mode T...

Страница 4: ...ation To prevent accidental loss of data the timer must be reset before the mode menu can be entered If single timer operation is selected the timer will give a prompt for the desired number of split...

Страница 5: ...external input is the only way to activate run1 count1 3 Stop2 Run2 In single timer mode this input is for stop2 in multiple timer mode it is for run count2 This external input is the only way to act...

Страница 6: ...ries is provided on the bottom of the 54035A Using a screwdriver or coin turn each of the thumbscrews and remove the battery cover plate The batteries are provided with two ribbon loops to aid in remo...

Страница 7: ...sends and receives only ASCII characters Each control is activated by sending a single ASCII character the timer is case sensitive ASCII Function S Start 1 Stop1 2 Stop2 3 Stop3 4 Stop4 C Begin run c...

Страница 8: ...1 Selecting the comm port in use by the mouse will result in an error and or disabling of the mouse If the mouse is disabled the computer will have to be restarted to restore function Timing function...

Страница 9: ...g results start and stop the timer from the front or back panel on the 54035A and use the computer to upload the time data Results File Format All time and count data is stored in an ASCII format Any...

Страница 10: ...D input on the timer The54035Awilldisplaythetimeontargetonthemaintimer timeofftargetontimer1andtherevolutioncount on timer 2 press scroll in multiple timer mode to view timer 2 data If the rotary purs...

Страница 11: ...binding posts on the Stability Platform to an input and ground on the 54035A 2 Connect the center binding post to the RUN CNT input 3 Connect the right binding post to RUN1 input 4 Connect the left bi...

Страница 12: ...nding posts on the Stability Platform to an input and ground on the 54035A 2 Connect the center binding post to the RUN CNT input 3 Connect the right binding post to RUN1 input 4 Connect the left bind...

Страница 13: ...rom the maze to the START input on the Multifunction Timer Counter 3 Connect the stop jack from the maze to STOP3 on the timer 4 Connect the error jack on the maze to RUN1 on the timer 5 Connect the s...

Страница 14: ...t of each switch to the one of the four GND inputs on the rear of the 54035A To start an event a remote pushbutton can be connected to the START input on the rear of the 54035A or a trial can begin by...

Страница 15: ...ing post on the board connects to any GND input on the 54035A The stylus for the tapping board connects to the RUN CNT input The 54035A will show the number of counts or taps along with the contact ti...

Страница 16: ...ranty repairs Damaged Goods Damaged instrumentation should not be returned to Lafayette Instrument prior to a thorough inspection If a shipment arrives damaged note damage on delivery bill and have th...

Отзывы: