Code 5: Advance drive overload.
If the blade has collided with an obstruction, restart the instrument and carefully
drive the blade away from the obstruction.
If the error occurs repeatedly during normal use contact Technical Support for
advice.
Code 6: Oscillating head fault. Restart the instrument. If the problem persists contact
Technical Support for advice.
Code 7: Oscillating head fault.
If motion of the vibrating head has been obstructed, restart the instrument and
carefully drive the blade away from the obstruction.
If the error occurs repeatedly during normal use contact Technical Support for
advice.
Code 8: Instrument overheat fault. Contact Technical Support for advice.
Code 9: Memory Fault. Restart the instrument. If the problem persists contact
Technical Support for advice.
Code 10: Alignment tool communications error.
If the alignment tool has been unplugged during the blade alignment
procedure, restart the instrument.
If the error occurs repeatedly during normal use contact Technical Support for
advice.