Conversely, the blade holder can be tilted in a counter-clockwise direction by turning
the adjusting screw counter-clockwise.
Press ‘SLICE’ to set the blade vibrating again and note the Z axis deflection. Stop the
blade vibrating. If the error has increased you should repeat the adjustment but in the
opposite direction.
Continue the above process until the Z axis deflection has been reduced to
acceptable limits. Once the blade is within acceptable limits the lower locking screw
may be fully tightened.
Recheck the blade alignment. Repeat the adjustment procedure if required.
5000mz
Loosen the upper and lower blade holder securing screws and retighten them just
sufficiently to grip the blade holder without locking it in place.
The blade holder can be rotated clockwise by first loosening the right hand set screw
and then tightening the left hand set screw.
Conversely, the blade holder can be rotated counter-clockwise by first loosening the
left hand set screw and then tightening the right hand screw.
Press ‘SLICE’ to set the blade vibrating again and note the Z axis deflection. Stop the
blade vibrating. If the error has increased you should repeat the adjustment but in the
opposite direction.
Continue the above process until the Z axis deflection has been reduced to
acceptable limits. Once the blade is within acceptable limits the cap head securing
screws may be fully tightened.
Recheck the blade alignment. Repeat the adjustment procedure if required.
7000smz
and
5000mz
When you have completed the alignment procedure press the ‘RETURN’ key and
follow the removal instructions as shown on the display.
In certain circumstances, for example if a steel blade is fitted and you have specified
a ceramic blade, the control system may position the blade incorrectly and not detect
the blade edge. In these instances the display will show the following message:
Correct any problems and press ‘SLICE’ to retry.