Model 54060A Instruction Manual
www.lafayetteinstrument.com . [email protected]
3
Wiring Instructions
24 AWG min
16 AWG max
Stripped portion length of wire should be a minimum
of 8mm and a maximum of 12mm
Insertion:
Depress the white push button with
screwdriver (shown at right), and place stripped wire
end into corresponding terminal block hole
Removal:
Depress the white push button; pull wire out
of hole
Buttons
Menu Button:
Pressing the Menu button while no test
is running will cycle through the following options.
Press either the Reset button or Pause button to select
the first or second option respectively. Each time
the device is powered on, the bolded option is set by
default.
•
RUN:
NO
NC
•
START:
NO
NC
•
STOP:
NO
NC
•
RESET:
NO
NC
•
TONE:
OFF
ON
•
START:
SNGL
REP
Reset Button:
Pressing the Reset button is identical to activating the Reset input. The Reset
input is invalid if the timer is running or if the device is paused.
Pause Button:
Pressing the Pause button toggles the Pause state. During the Pause state,
no inputs can be activated and the only button that works is the Pause button to toggle out
of the Pause state.