Replacing and Filling IO Units with Blood
The IO Pads supplied with SimMan 3G Trauma are disposable parts meant for single use.
Sternal IO
1 Attach the Sternal IO bag to the Sternal tube and close
off the pinch clamp.
2 Remove the Sternal IO pad from the manikin chest.
3 Remove the Sternal tube from the sternal pad.
4 Fill the sternal IO pad with 7ml of blood. Ensure the pad
is completely filled.
5 Reattach the Sternal tube to the Sternal IO pad.
6 Place the Sternal IO unit in the Sternal Chassis. Slide
the unit down and towards the neck, until it fits securely
under the top edge of the chassis.
7 Reattach the sternal IO pad
The sternal IO is now ready for simulation.
Note: If bits of plastic from the module get stuck in the needle, flush the needle with fluid to dislodge the obstruction.