Replacing veins: (Keeping skin)
1 . Lubricate inside of skin with liquid detergent; let it flow down
into the finger area .
2 . Begin at top of arm and slowly pull skin down and off of arm .
Do not roll, as that will cause skin to bind .
3 . Remove tubing from track in mandrel . Glue may need
to be scraped away to allow removal .
4 . Rinse and dry vein grooves well and swab with alcohol .
Be sure to remove any excess glue .
5 . Place new veins along grooves, (Figure 5) spot gluing
as needed . (We recommend a fast-drying glue .)
6 . Generously lather arm mandrel with liquid soap .
Heat arms skins before replacing on mandrel .
This can be done with a blow-dryer .
7 . Slide hand into skin .
8 . Work skin over fingers, as with a glove .
9 . Work arm skin up, over mandrel .
SimPad Connection
1 . Connect Nursing Anne to
Link Box, via cable located on
back of manikin . (Photo 3) .
For some manikins the
adapter cable supplied with
SimPad may have to be used .
2 . Connect clear tubing exiting
manikin together with the
cable .
See SimPad DFU for complete
Heart, Breath and Bowel Sound
Auscultation.
Blood Pressure Arm –
1 . Place Blood Pressure Cuff
on arm .
2 . Attach tubing on cuff
to clear tube,
located underneath
Nursing Anne’s arm .
(Photo 4)
See SimPad DFU for complete
Blood Pressure operating procedures.
Do not insert needles into blood pressure arm .
Auscultation of Heart, Breath and Bowel Sounds and
Fetal Tones – See SimPad DFU for complete operating procedures.
Genitalia:
Both male and female genitalia have been provided for urinary
catheterization and enema training procedures .
Tube-like valves, with a screw cap on one end, connect urethra and anus
to corresponding reservoirs . This cap must be connected to genitalia
openings . Soft vinyl end of connector is attached to urinary and anal
reservoirs with a white clamp . (Long valve connects urinary reservoir;
short valve connects anal reservoir .)
Pelvic support pin must
be removed and replaced
immediately when assembling
and disassembling genitalia .
(Figure 6)
Pelvic support pin must be replaced.
Failure to replace pin will result in pelvic spread .
If this occurs, the genitalia module will no longer fit
properly in the manikin .
To disassemble:
1 . Lift up on belly plate .
2 . Pull upper portion of genitalia back to remove pelvic
support pin .
3 . Belly plate, genitalia with connectors, and colon reservoir
may now be removed simultaneously .
To assemble:
(Photos 5, 6 & 7)
1 . Screw both valves onto genitalia . When attaching valves, be sure
alignment of genitalia, urinary reservoir (belly plate) and colon
reservoir are as illustrated in Figure 7 .
2 . Lift up on edge of belly plate .
3 . Pull down upper portion of genital .
4 . Insert into pelvic cavity .
5 . Replace the pelvic support pin .
6 . Press belly plate into position .
Long valve connects urinary reservoir;
short valve connects anal reservoir .
Figure 5
Photo 3
Photo 4
Figure 7
Photo 5
Slide tube to meet ridge
on connector .
Photo 6
Area where clamp
is placed .
Photo 7
Apply clamp and squeeze
to tighten .
Figure 6
Содержание Nursing Anne
Страница 2: ......
Страница 59: ...Nursing Anne 3 Laerdal a b c a b 4 5 3 1 1 1 2 3 4 5 6 7 1 8 6 1 1 1 1 1 2 3 4 5 6 7 2 3 22 1 2 3 4 3 4 2 3 4 1...
Страница 61: ...Nursing Anne 5 Laerdal 8 8 a b c d 16 French 7 C 1 2 3 4 5 6 8 7 1 2 3 4 5 6 7 8 9 9 9 10 3 11 8 8 9 9...
Страница 67: ...Nursing Anne 5 Laerdal 8 8 a O b c d 16Fr 7mm C 1 2 3 4 5 6 8 7 1 2 3 4 5 6 7 8 9 9 9 10 3 11 8 9 8 9...
Страница 75: ......