laerdal Bemis 78 00 20 Скачать руководство пользователя страница 4

2.1 Unpacking

-  Remove the LSU from its packaging and inspect all parts for 
  damage. If there are any signs of damage or if any parts are missing, 

immediately notify Laerdal or one of its authorized dealers. 

-  Insert the battery according to section 5.6. 
-  Place the LSU on charge for a minimum of 24 hours to make sure you have 

a fully charged battery (see section 5.4).

-  Perform the Device Test according to section 3.2.3.

2.2 Indications for Use

The LSU is a portable, electrically powered, medical suction equipment 
intended for field and transport use. It is intended for intermittent 
operation to remove secretions, blood or vomit from a patient´s airway
to allow ventilation. Higher vacuum levels are generally selected for 
oropharyngeal suctioning, and lower vacuum levels are usually selected 
for tracheal suctioning and the suctioning of children and infants.

2.3 Contraindications

None known.

2.4 Features and Benefits 

-  Splashproof, rugged and solid.
-  Easy access to the tubing.
-  Equipped with an internal battery that can easily be replaced 
  without using tools.   
-  Charging of battery can be done directly from AC or DC mains - no 

external charger is needed. 

-  Fast charge option of 3 hours, giving approx. 80% battery capacity.  
-  Operation possible from internal battery and directly from AC or DC mains 

- no adapter is needed.

-  A large combined ON/OFF switch and vacuum selector makes 
  operation easy even with heavy-duty gloves.  
-  A green LED bargraph makes it easy to observe the actual vacuum level 

during operation.

-  Another green LED bargraph indicates approx. remaining battery capacity.

4

2 Introduction

LSU DfU US 6791 rev H.indd   4

30.09.2016   14:39:49

Содержание Bemis 78 00 20

Страница 1: ...Laerdal Suction Unit Directions for Use Cat No 78 00 20 Bemis www laerdal com LSU DfU US 6791 rev H indd 1 30 09 2016 14 39 49...

Страница 2: ...ce the LSU on continuous charge when not in use make sure the battery is charged for a minimum of 24 hours at least once a month Do not store the battery when it is discharged Always fully charge the...

Страница 3: ...ter 12 3 4 Power Sources Connector Sockets 13 3 5 Automatic Power save Function 13 4 Operation 14 4 1 Getting Started 14 4 2 Replace the Canister 15 4 3 After Use 16 5 Maintenance 16 5 1 General 16 5...

Страница 4: ...r oropharyngeal suctioning and lower vacuum levels are usually selected for tracheal suctioning and the suctioning of children and infants 2 3 Contraindications None known 2 4 Features and Benefits Sp...

Страница 5: ...ive 5 year limited warranty See the enclosed Laerdal Global Warranty for terms and conditions The warranty is also available at www laerdal com Excluding the canister tubing systems and battery LSU Df...

Страница 6: ...nnector Patient Suction Tubing Canister Holder Battery Door Canister Holder Release Arm Canister Canister Lid w Bacterial Filter Shut Off Valve Strap for Patient Suction Tubing Vacuum Connector Tubing...

Страница 7: ...350 500 mmHg The free air flow value changes along with the increasing vacuum settings see section 9 4 for details The LSU is switched OFF by resetting the Operating Knob to 0 Even if the Operating K...

Страница 8: ...s to be lit after three OFF ON cycles and after replacing the battery with a fully charged battery discontinue use and return the LSU to service See section 6 Failure Mode Indication caused by externa...

Страница 9: ...LEDs lit 20 1 LED lit Battery Low min 1 minute of operation left based on 500 mmHg free flow 1 LED flashes Note If the LSU or the NiMH battery has been stored at low temperatures 12 C 54 F the LSU ma...

Страница 10: ...E TEST Before you press the TEST Button make sure the Patient Suction Tubing is not occluded or bent DEVICE TEST DESCRIPTION Press and hold the TEST Button while setting the Operating Knob to 500 mmHg...

Страница 11: ...DISPLAYED ON THE BATTERY STATUS INDICATOR STEP 1 Occlusions STEP 2 Vacuum Build Up Efficacy within 3 seconds STEP 3 Max Achievable Vacuum within 10 seconds STEP 4 Air Leakage TEST RESULT DISPLAYED ON...

Страница 12: ...contains a Shut Off Valve that shuts off the vacuum if the Canister is full or the LSU tips over To reset the valve remove the Angled Connector from the VACUUM Inlet Note All parts are disposable and...

Страница 13: ...ilable separately see section 8 External power connection will cause the External Power Indicator to be lit 3 5 Automatic Power save Function The LSU has an Automatic Power save Function that switches...

Страница 14: ...o the PATIENT Inlet on the Canister Lid 4 Check that a suction catheter is attached to the patient suction tube or suction adapter Do not use the suction tube or suction adapter without a suction cath...

Страница 15: ...follows 2 Disconnect the Angled Connector from the VACUUM Inlet on the Lid Connect the suction end of the Patient Suction Tubing to the VACUUM Inlet to avoid spillage 3 To release the Canister Holder...

Страница 16: ...attery Quality Check see section 5 5 for details or after 3 years Do not store the LSU with an discharged battery 5 3 Clean the Cabinet Warning Disconnect the LSU from external power prior to cleaning...

Страница 17: ...ended to always fully charge the battery Repetitive charging to a lower capacity level will reduce battery life If it is not possible to place the LSU on continuous charge when not in use make sure th...

Страница 18: ...e LSU on charge unless a fully charged battery is inserted There are no user serviceable parts inside the cabinet Do not open the cabinet Refer servicing to personnel qualified by Laerdal or to Laerda...

Страница 19: ...nnection between Pump Unit and Canister Patient Suction Tubing twisted or blocked Battery is not charged Tube s is kinked or twisted ACTION Check power cord connec tions and the external AC or DC powe...

Страница 20: ...wer Cord not illustrated 78 23 00 Shoulder Strap 78 24 00 Side Pouch 78 04 40 External charger kit not illustrated Parts 78 03 00 Direction for Use not illustrated 78 04 20 Bemis Canister Holder 78 04...

Страница 21: ...y the MEDICAL ELECTRICAL EQUIPMENT should be observed to verify normal operation in the configuration in which it will be used 9 2 General tolerance Overall tolerance 5 9 3 Physical Characteristics Si...

Страница 22: ...e Butadiene Styrene PC ABS Protector for front Styrene Ethylene Betyl Styrene SEBS Cabinet back PC ABS Cabinet base w Protector PC ABS SEBS Battery door SEBS Connector retainer for battery Poly Oxy Me...

Страница 23: ...ective 2012 19 EC on Waste Electrical and Electronic Equipment WEEE The symbol on the product or on the documents accompanying the product indicates that this appliance may not be treated as household...

Страница 24: ...ns power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2kV common mode 1 kV differential mode 2kV common mode Mains power quality sh...

Страница 25: ...as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symb...

Страница 26: ...SU uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The L...

Страница 27: ...LSU DfU US 6791 rev H indd 27 30 09 2016 14 39 53...

Страница 28: ...Distributed in USA by LAERDAL MEDICAL CORPORATION 167 Myers Corners Road P O Box 1840 Wappingers Falls NewYork 12590 8840 Tel 1 800 431 1055 1 845 297 7770 Fax 1 800 227 1143 1 845 298 4545 2016 Laerd...

Отзывы: