background image

Exit from calibration by pressing button 

 

.

CONFIGURATION PARAMETERS

To get access to the parameter configuration menu, press button  

 

 +    for 5 seconds.

With button  

 

 or  

  select the parameter to be modified.

Press button  

 

 to display the value.

By keeping button  

 

 pressed, use button 

 or    to set the desired value. 

When button  

 

 is released, the newly programmed value is stored and the following parameter is displayed.

To exit from the setup, press button 

 

 or wait for 30 seconds.

PAR

RANGE DESCRIPTION

SCL

1°C;

2°C;

°F

Readout scale (see table of input specifications)

Caution: upon changing the 

SCL

 value, it is then absolutely necessary to reconfigure the param-

eters relevant to the absolute and relative temperatures (

SPL

SPH

1SP

1HY

  etc..)

SPL

-50°...SPH

Minimum limit for 

1SP

 setting

SPH

SPL...150°

Maximum limit for 

1SP

 setting.

1SP

SPL... SPH Setpoint (value to be maintained in the room).

1CM

HY; PID

Control mode.

With 

1CM

=HY you select control with hysteresis: parameters 

1HY

1T0

 and 

1T1

 are used.

With 

1CM

=PID you select a Proportional-Integral-Derivative control mode: parameters 

1PB

1IT

1DT

1AR

1CT

 will be used

1CH

REF; HEA

Refrigerating (REF) or Heating (HEA) control mode.

1CM

=HY

1HY

0...19.9°

OFF/ON thermostat differential.  With 

1HY

=0 the output is always off. 

  

1T0

0...30min

Minimum off time.

After  output  1  has  been  turned  off,  it    remains  inactive  for 

1T0

  minutes  regardless  of  the 

temperature value measured.

1T1

0...30min

Minimum on time. (

the following parameter will be 

1PF

).

After output 1 has been turned on,  it remains active for 

1T1

 minutes regardless of the temperature 

value measured.

1CM

=PID

1PB

0...19.9°

Proportional bandwidth.   

Temperature control takes place by changing the 

ON time of the output: the closer the temperature 

to the setpoint, the less time of activation. A small 

proportional  band  increases  the  promptness  of 

response of the system to temperature variations, 

but  tends  to  make  it  less  stable.  A  purely 

proportional  control  stabilises  the  temperature 

within  the  proportional  band  but  does  not  cancel 

the deviation from setpoint. 

With 

1PB

=0 the output is always off.

1IT

0...999s

Integral action time.    

The steady-state error is cancelled by inserting an 

integral action. The integral action time, determines 

the speed with which the steady-state temperature 

is achieved, but a high speed (

1IT

 low) may be the 

cause of overshoot and instability in the response. 

With 

1IT

=0 the integral control is disabled.

1DT

0...999s

Derivative action time.   

Response  overshoot  may  be  reduced  by  inserting 

a  derivative  Action.  A  high  derivative  action  (

1DT

 

high)  makes  the  system  very  sensitive  to  small 

temperature  variations  and  causes  instability.  With 

1DT

=0 the derivative control is disabled.

1AR

0...100%

Reset of integral action time referred to 

1PB

Decreasing the parameter 

1AR

 reduces the integral control action zone, and consequently the 

overshoot (see figure on paragraph 

1IT

).

1CT

1...255s

Cycle time. 

It’s the period in which the output ON time changes. The quicker the system to be controlled 

reacts to temperature variations, the smaller the cycle time must be, in order to obtain higher 

temperature stability and less sensitivity to load variations.

1PF

ON/OFF

Output state in case of probe failure.

OAU

NON;

THR;

AL0;

AL1

AUX output operation.

NON : output disabled (always off). (

the next parameter will be 

ATM

)

THR: output programmed for second thermostat control (

the next parameter will be 

2SM

).

AL0: contacts open when an alarm condition occurs (

the next parameter will be 

ATM

).

AL1: contacts make when an alarm condition occurs (

the next parameter will be 

ATM

).

OAU

=THR

2SM

ABS;

REL

Setpoint 2 mode.

Channel 2 setpoint may be absolute (

2SM

=ABS), or a differential relative to setpoint 1 (

2SM

=REL)

2SM

=ABS

2SP

SPL...SPH Auxiliary output switchover temperature (

the next parameter will be 

2CH

)

 

2SM

=REL

2DF

-19.9...19.9°

Temperature differential relative to 

1SP

. The auxiliary output setpoint is equal to 

1SP

+

2DF

     

ON/OFF refrigerating control 

(

1CM

=HY, 

1CH

=REF)

ON/OFF heating control 

(

1CM

=HY, 

1CH

=HEA)

1SP 1SP+1HY

T[°]

ON

OFF

1SP

1SP-1HY

T[°]

ON

OFF

Time

Temperature

Process

temperature

Overshoot

Steady-state error

1PB

1SP

Time

Temperature

Process

temperature

Overshoot

1PB

1PBx1AR%

integral control

action area

1SP

Time

Temperature

Process

temperature

Overshoot

1PB

1SP

2SP 2SP+2HY T[°]

ON

OFF

2SP

2SP-2HY

T[°]

ON

OFF

ON/OFF control in refrigeration

(

2SM

=ABS, 

2CH

=REF)

ON/OFF control in heating

(

2SM

=ABS, 

2CH

=HEA)

1SP+2DF

1SP

2DF>0

1SP+2DF+2HY

T[°]

ON

OFF

1SP

1SP+2DF

2DF<0

1SP+2DF-2HY

T[°]

ON

OFF

ON/OFF control in refrigeration. Setpoint 2

relative to setpoint 1 (

OAU

=THR, 

2CH

=REF)

ON/OFF control in heating. Setpoint 2

relative to setpoint 1 (

OAU

=THR, 

2CH

=HEA)

TECHNICAL DATA

Power supply

AC1-5…D  12Vac/dc ±10%, 2W

AC1-5...W  110 - 230Vac±10%, 50/60Hz, 2W

Relay outputs (AC1-5..R..)

OUT1 

12(4)A

OUT2 

7(2)A

SSR drive (AC1-5..M..)

OUT1 

15mA   12Vdc

Inputs

see table of input specifications

Measurement range

see table of input specifications

Measurement accuracy

see table of input specifications

Operating conditions

-10 … +50°C; 15%...80% U.R.

CE (Reference Norms)

EN60730-1; EN60730-2-9;

EN55022 (Class B); EN50082-1

Front protection

IP55

VIA PADOVA, 25

31046 ODERZO /TV /ITALY

TEL. +39 - 0422 815320

FAX +39 - 0422 814073

www.lae-electronic.com

E-mail: [email protected]

OAU

=THR

2CH

REF; HEA

Refrigerating control (REF) or heating control mode (HEA) for the auxiliary output.

2HY

0...19.9°

Differential of thermostat 2. With 

2HY

=0 the auxiliary output always remains off.

2T0

0...30min

Minimum off time.

After output 2 has been turned off, it remains inactive for 

2T0

 minutes regardless of the temperature 

value measured.

2T1

0...30min

Minimum on time.

After output 2 has been turned on, it remains active for 

2T1

 minutes regardless of the temperature 

value measured.

2PF

ON/OFF

Auxiliary output state in case of probe failure.

ATM

NON;

ABS;

REL

Alarm threshold management.

NON: all temperature alarms are inhibited (

the following parameter will be 

SB

).

ABS: the values programmed in 

ALA

 and 

AHA

 represent the real alarm thresholds.

REL: the values programmed in 

ALR

 and 

AHR

 are alarm differentials referred to 

1SP

 and 

1SP

+

1HY

.

ATM

=ABS

ALA

-50°...AHA Low temperature alarm threshold.

AHA

ALA...150° High temperature alarm threshold.

ATM

=REL

ALR

-12.0...0°

Low temperature alarm differential. 

With 

ALR

=0 the low temperature alarm is excluded

AHR

0...12.0°

High temperature alarm differential. 

With 

AHR

=0 the high temperature alarm is excluded

ATD

0...120min

Delay before alarm temperature warning.

SB

NO/YES

Stand-by button enabling.

INP

0mA/4mA,

T1/T2

ST1/SN4

Sensor input selection (see table of input specifications).

In the models AC1-5A..., AC1-5J..., AC1-5T... only.

RLO

-19.9...RHI

Minimum range value (

in the models AC1-5A…, AC1-5I… only

)

RLO

 takes the minimum value measured by the transmitter (i.e. the value matching 0V, 0/4mA).

RHI

RLO...99.9

Maximum range value (

in the models AC1-5A…, AC1-5I… only

)

RHI

 takes the maximum value measured by the transmitter (i.e. the value matching 1V, 20mA)

OS1

-12.5...12.5° Probe T1 offset.

TLD

1...30min

Delay for minimum temperature (TLO) and maximum temperature (THI) logging.

SIM

0...100

Display slowdown

ADR

1...255

AC1-5 address for PC communication

INPUT SPECIFICATIONS

MODEL

INPUT

RANGE [MEASUREMENT ACCURACY]

SCL=1°C

SCL=2°C

SCL=°F

AC1-5A...

0÷1V

RLO÷RHI [< ± 3mV]

---

AC1-5I...

INP = 0mA

0÷20mA

RLO÷RHI [< ± 0.2mA]

---

INP = 4mA

4÷20mA

AC1-5J...

INP=T1

TC “J”

---

-50÷750°C [ < ±3°C ]

-60÷999°F [ < ±5°F ]

INP=T2

TC “K”

---

-50÷999°C [ < ±3°C ]

AC1-5P...

PT100

-50/-19.9÷99.9/150°C

[ < ±0.3°C ]

-100÷850°C

[ <±1°C(-50÷850°), ±2°C ]

-150÷999°F

[ <±2°F(-60÷999°), ±4°F]

AC1-5T...

INP=ST1

PTC 1000 Ω

(LAE ST1..)

-50/-19.9 ÷ 99.9/150°C

[<±0.3°C(-30÷130°),±1°C]

-50 ÷ 150°C 

[<±0.3°C(-30÷130°), ±1°C]

-60 ÷ 300°F 

[< ±0.6°F(-20÷260°),±2°F]

INP=SN4

NTC 10K Ω 

(LAE SN4..)

-40/-19.9 ÷ 99.9/125°C

[<±0.3°C(-40÷100°),±1°C]

-40 ÷ 125°C

[<±0.3°C(-40÷100°),±1°C]

-40 ÷ 260°F

[<±0.6°F(-40÷210°), ±2°F]

WIRING DIAGRAMS

 

 

 

 

 

 

1SP

1SP-ALR

T[°]

ON

OFF

1SP+1HY+AHR

1SP

1SP-1HY-ALR

1SP+AHR

T[°]

ON

OFF

Temperature alarm with relative thresholds, 

refrigerating control (

ATM

=REL, 

1CH

=REF)

Temperature alarm with relative thresholds, 

heating control (

ATM

=REL, 

1CH

=HEA).

AC1-5AS2RW-B

data I/O

115...

~

230V

OUT2

RS485

7(2)A

OUT1

12(4)A

rH

V

9 10 11

V

IN

V+

V-

1

2

3

4

5

6

7

AC1-5JS2RW-A

data I/O

115...

~

230V

OUT2

TTL

7(2)A

OUT1

12(4)A

+

-

10

11

1

2

3

4

5

6

7

AC1-5PS2MW-B

data I/O

115...

~

230V

OUT2

RS485

7(2)A

1

2

12V

OUT1

15mA

SSR

+

-

W R W

9 10 11

1

2

4

5

6

7

AC1-5TS2RW-A

data I/O

115...

~

230V

OUT2

TTL

7(2)A

OUT1

12(4)A

10

11

1

2

3

4

5

6

7

DISPLAY

During normal operation, the display shows either the temperature measured or one of the following indications:

OFF

Controller in stand-by

TUN/xx.x

Controller in autotuning

OR

Probe T1 overrange or failure

E1

In tuning: timeout1 error

HI

Room high temperature alarm

E2

In tuning: timeout2 error

LO

Room low temperature alarm 

E3

In tuning: overrange error

MENU INFO

The information available in this menu is:

THI

Maximum temperature recorded

LOC

Keypad state lock

TLO

Minimum temperature recorded

Access to menu and information displayed.

Press and immediately release button 

 

With button 

 

 or    select the data to be displayed.

Press button 

 

 to display value.

To exit from the menu, press button 

 

 or wait for 10 seconds.

Reset of THI, TLO recordings

With button 

 

 or   select the data to be reset.

Display the value with button 

 

.

While keeping button 

 

 pressed, use button 

CHANNEL 1 SETPOINT 

(display and modification of desired temperature value)

 

Press and release button 

 

 : the LED L1 blinks, the display shows 1SP for 1 second and then the setpoint associated value.

Press buttons 

 

 or 

 to set the desired value (adjustment is within the minimum 

SPL

 and maximum 

SPH

 limit).

To store the new value press button 

 

, or wait for 10 seconds.

To go back to normal mode without saving the new value, press 

 

CHANNEL 2 SETPOINT

With the auxiliary output set as thermostat control (

 

OAU

=THR), it’s possible to modify setpoint 2 during the normal operation 

of the controller.

Press and release button 

 

: the LED L2 blinks, the display shows 2SP for 1 second if setpoint 2 is an absolute threshold 

(

2SM

=ABS), alternatively the display shows 2DF, if setpoint 2 is a threshold relative to setpoint 1 (

2SM

=REL), then the value 

associated to the parameter appears.

Press buttons 

 

 or   to set the desired value.

To store the new value press button 

 

 or wait for 10 seconds.

To go back to normal mode without saving the new value, press 

 

.

STAND-BY

Button 

, when pressed for 3 seconds, allows the controller to be put on a standby or output control to be resumed (with 

SB

=YES only).

KEYPAD LOCK

The keypad lock avoids undesired, potentially dangerous operations, which might be attempted when the controllers is operating 

in a public place. In the INFO menu, set parameter 

LOC

=YES to inhibit all functions of the buttons. To resume normal operation 

of keypad, adjust setting so that 

LOC

=NO.

CONTROLLER AUTOTUNING IN PID MODE

Before starting

In  the  setup  mode  (see  configuration  parameters):  set 

1CM

=PID;  make  sure  that 

1CH

  matches  the  desired  operation  mode 

(

1CH

=REF for refrigerating control, 

1CH

=HEA for heating control); then adjust setpoint 

1SP

 at the desired value.

Start autotuning

During normal operation, keep buttons 

 + 

 pressed for 3 seconds. 1CT blinks on the display. With 

 +   or   set the cycle 

time in order to define the dynamic of the process to be controlled. To abort the autotuning function, press 

; to start autotuning 

press   + 

 or wait for 30 seconds.

During autotuning

During the entire autotuning phase, the display alternates TUN with the actual temperature measured. In case of power failure, 

when power is resumed, after the initial autotest phase, the controller resumes the autotuning function. To abort the autotuning, 

without modifying the previous control parameters, keep button   pressed for 3 seconds. After the autotuning has taken place 

successfully, the controller updates the control parameters and start to control.

Errors

If the autotuning function failed, the display shows an error code:

E1 timeout1 error: the controller could not bring the temperature within the proportional band. Increase 

 

1SP

 in case of heating 

control, vice versa, decrease 

1SP

 in case of refrigerating control and re-start the process.

E2 timeout2 error: the autotuning has not ended within the maximum time allowed (1000 cycle times). Re-start the autotuning 

 

process and set a longer cycle time 

1CT

.

E3 temperature overrange: check that the error was not caused by a probe malfunction, then decrease 

 

1SP

 in case of heating 

control, vice versa increase 

1SP

 in case of refrigerating control and then re-start the process.

To eliminate the error indication and return to the normal mode, press button 

 

.

Control improvement

To reduce overshoot, reduce the integral action reset 

 

1AR

To increase the response speed of the system, reduce the proportional band 

 

1PB

. Caution: doing this makes the system less stable.

To reduce swings in steady-state temperature, increase the integral action time 

 

1IT

; system stability is thus increased, although 

its response speed is decreased.

To increase the speed of response to the variations in temperature, increase the derivative action time 

 

1DT

. Caution: a high 

value makes the system sensitive to small variations and it may be a source of instability.

RECALIBRATION

Have a precision reference thermometer or a calibrator to hand. Ensure that 

 

OS1

=0 and 

SIM

=0.

Switch the controller off then on again.

 

During the auto-test phase, press buttons 

 

   

  and keep them pressed till the controller shows 

0AD

.

With buttons 

 

 and    select 

0AD

 or 

SAD

0AD

 allows a calibration of 0, inserting a constant correction over the whole scale 

of measurement. 

SAD

 allows a calibration of the top part of the measurement scale with a proportional correction between the 

calibration point and 0.

Press 

 

 to display the value and then use    +   or 

  to make the read value coincide with the value measured by the 

reference instrument. 

OPERATION

Insert the controller through a hole measuring 71x29 mm;

 

Make sure that electrical connections comply with the paragraph “wiring diagrams”. To reduce the effects of electromagnetic 

 

disturbance, keep the sensor and signal cables well separate from the power wires. 

Fix the controller to the panel by means of the suitable clips, by pressingly gently; if fitted, check that the rubber gasket adheres 

 

to the panel perfectly, in order to prevent debris and moisture infiltration to the back of the instrument.

ATTENTION: during the setup of the controller, please make sure that the parameter INP matches the sensor used, as indicated 

 

in the table “input specifications”.

Place the probe T1 inside the room in a point that truly represents the temperature of the stored product.

 

INSTALLATION

OUT1

OUT2

L1

L2

Channel 1 output
Channel 2 output

Alarm

Channel 1 setpoint modification
Channel 2 setpoint modification

Info / Enter button

Fig.1 - Front panel

Modify Setpoint 1 / Decrease button

Increase / Modify Setpoint 2 button
Exit / Stand-by button.

Thank you for having chosen a LAE electronic product. Before installing the instrument, please read these instructions carefully

to ensure maximum performance and safety.

DESCRIPTION

INDICATION

AC1-5 INSTRUCTION FOR USE

Содержание LTR-5 Series

Страница 1: ...Lots of Options just One Choice...

Страница 2: ...L ots of O ptions just O ne C hoice...

Страница 3: ...supply no serial port On request the LTR 5 is also available with gasket for a better protection between bezel and panel a 50 140 C b 40 110 C c remaining range LTR 5 Series POS FUNCTION 1 2 3 4 5 1 L...

Страница 4: ...6 7 230V data I O OUT1 S S R 4 12V 3 10 11 RS485 LTR 5CSFE B 6 7 230V data I O 2 3 4 OUT1 16 4 A 10 11 RS485 LTR 5TSRE B Wiring diagrams LTR 5...

Страница 5: ...ture Control of small cold stores refrigerated cabinets and tables heating systems heated cupboards bains marie ovens laboratory equipment Humidity Control of greenhouses seasoning cells cold rooms ai...

Страница 6: ...d value When button is released the newly programmed value is stored and the following parameter is displayed To exit from the setup press button X or wait for 30 seconds PAR RANGE DESCRIPTION SCL 1 C...

Страница 7: ...age on the outputs is 50Vac dc in order to ensure safety insulations AC1 5 Series AC1 5T Functions Range Input type Accuracy PTC 50 150 C 60 300 F 0 3 C NTC10K 40 125 C 40 260 F 0 3 C AC1 5A AC1 5I 0...

Страница 8: ...12 4 A TTL 7 2 A 115 230V OUT1 OUT2 10 11 1 2 3 4 5 6 7 AC1 5TS2RW A Wiring diagrams AC1 5 15mA RS485 7 2 A 115 230V OUT1 OUT2 10 9 11 1 2 4 5 6 7 W R W SSR 12V AC1 5PS2MW B...

Страница 9: ...ture Control of small cold stores refrigerated cabinets and tables heating systems heated cupboards bains marie ovens laboratory equipment Humidity Control of greenhouses seasoning cells cold rooms ai...

Страница 10: ...00 850 C 1 C 50 850 2 C 150 999 F 2 F 60 999 4 F AC1 5T INP ST1 PTC 1000 LAE ST1 50 19 9 99 9 150 C 0 3 C 30 130 1 C 50 150 C 0 3 C 30 130 1 C 60 300 F 0 6 F 20 260 2 F INP SN4 NTC 10K LAE SN4 40 19 9...

Страница 11: ...e Input type Accuracy PTC 50 150 C 60 300 F 0 3 C NTC10K 40 125 C 40 260 F 0 3 C AC1 27A AC1 27I 0 1V Configurable in setup 3mV 0 4 20mA 0 2mA AC1 27J TC J 50 750 C 60 999 F 3 C TC K 50 999 C 60 999 F...

Страница 12: ...I O 230V OUT1 16 4 A 2 3 4 V IN V V 21 20 19 18 17 16 12 11 RS485 OUT2 16 4 A rH V AC1 27AS2RE B Wiring diagrams AC1 27 data I O 230V OUT2 16 4 A OUT1 16 4 A 21 20 19 18 17 16 12 11 TTL 2 3 AC1 27JS2...

Страница 13: ...Temperature on control panels for small cold stores heating systems heated cupboards bains marie ovens laboratory equipment Humidity control panels for greenhouses seasoning cells cold rooms air condi...

Страница 14: ...NP 4mA 4 20mA AC1 27J INP T1 TC J 50 750 C 3 C 60 999 F 5 F INP T2 TC K 50 999 C 3 C AC1 27P PT100 50 19 9 99 9 150 C 0 3 C 100 850 C 1 C 50 850 2 C 150 999 F 2 F 60 999 4 F AC1 27T INP ST1 PTC 1000 L...

Страница 15: ...0 1V input detachable screw terminals 2 relays 12Vac dc supply voltage TTL port AC1 2W Series AC1 2WT AC1 2WA Functions Range Input type Accuracy PTC 50 150 C 60 300 F 0 3 C NTC10K 40 125 C 40 260 F...

Страница 16: ...data I O 230V OUT2 8 3 A OUT1 8 3 A 3 5 4 8 9 10 6 7 TTL rH V V IN V V AC1 2WAQ2RE A Wiring diagrams AC1 2W data I O 230V OUT2 8 3 A OUT1 8 3 A 3 4 8 9 10 6 7 RS485 AC1 2WTQ2RE B...

Страница 17: ...y slowdown ADR 1 255 AC1 2W address for PC communication INPUT SPECIFICATIONS MODEL INPUT RANGE MEASUREMENT ACCURACY SCL 1 C SCL 2 C SCL F AC1 2WA 0 1V RLO RHI 3mV AC1 2WT INP ST1 PTC 1000 LAE ST1 50...

Страница 18: ...L ots of O ptions just O ne C hoice...

Страница 19: ...els heated cabinets Applications AT1 5 Series AS1E G BS2E BG BS6E AL thermostat evaporator Functions Inputs door switch Outputs Power supply thermostat 16 4 A 230Vac TTL auxiliary 7 2 A thermostat 16...

Страница 20: ...iagrams AT1 5 1 2 3 5 4 9 8 7 10 11 16 5 A 7 2 A AUX TTL data I O PWR T1 T2 Door Switch Door Com AT1 5BS4E AG 1 2 3 5 4 9 8 7 10 11 16 8 A 7 2 A AUX RS485 data I O PWR T1 T2 Door Switch Door Com AT1 5...

Страница 21: ...Applications Freestanding upright cabinets and counters cold stores plug in display cases control panels heated cabinets...

Страница 22: ...relevant to the absolute and relative temperatures SPL SPH SP ALA AHA etc SPL 50 SPH Minimum limit for SP setting SPH SPL 120 Maximum limit for SP setting SP SPL SPH Setpoint value to be maintained in...

Страница 23: ...ries BS4E G BS4E AG BS4E AL thermostat evaporator Functions Inputs door switch Outputs Power supply thermostat 230Vac serial port TTL auxiliary evaporator fans Serial port generic Keypad with light bu...

Страница 24: ...TTL data I O 16 5 A 7 2 A 7 2 A AUX 12Vac dc 1 2 5 3 4 7 8 9 11 10 T1 T2 Door Switch Door Com 230Vac AT2 5BS4E AG...

Страница 25: ...Applications High or Low Temperature upright cabinets and counters cold stores plug in display cases control panels heated cabinets...

Страница 26: ...sary to reconfigure the parameters relevant to the absolute and relative temperatures SPL SPH SP ALA AHA etc SPL 50 SPH Minimum limit for SP setting SPH SPL 120 Maximum limit for SP setting SP SPL SPH...

Страница 27: ...refrigerated tables Applications Control Range 50 120 C 55 240 F Resolution 0 1 1 C F Accuracy NTC10K 0 3 C 40 0 70 0 C PTC1000 0 5 C 50 120 C Sensor type selectable NTC10K or PTC1000 Power supply 11...

Страница 28: ...AUX 7 2 A 7 2 A 12 5 A 8 3 A 5 6 115 230Vac RS485 data I O 7 8 AUX 17 18 DI2 Def sync 24 T3 Aux T2 T1 Door 3 19 21 20 22 23 2 1 4 AD2 5C34W BG...

Страница 29: ...Applications Plug in cabinets supermarket display cases cold stores control panels upright fridges and freezers refrigerated tables...

Страница 30: ...e the LAE models SN4 with INP ST1 the probes must be the LAE models ST1 OS1 12 5 12 5 C Probe T1 offset T2 NO YES Probe T2 enabling evaporator OS2 12 5 12 5 C Probe T2 offset T3 NON DSP CND 2EU Auxili...

Страница 31: ...ants where real time defrost starts are needed Applications Control Range 50 120 C 55 240 F Resolution 0 1 1 C F Accuracy NTC10K 0 3 C 40 0 70 0 C PTC1000 0 5 C 50 120 C Sensor type selectable NTC10K...

Страница 32: ...AUX 7 2 A 7 2 A 12 5 A 8 3 A 5 6 115 230Vac RS485 data I O 7 8 AUX 17 18 DI2 Def sync 24 T3 Aux T2 T1 Door 3 19 21 20 22 23 2 1 4 AR2 5C34W BG...

Страница 33: ...Applications Plug in cabinets supermarket display cases cold stores control panels upright fridges and freezers refrigerated tables and all those plants where real time defrost starts are needed...

Страница 34: ...ary output is switched on with a delay of 2CD seconds after the main compressor has cut in Both compressors are turned off at the same time INP SN4 ST1 Temperature sensor selection With INP SN4 the pr...

Страница 35: ...evaporator fan control Absolute or relative temperature alarms and door open alarm Temperature and pressure monitoring and condensing unit maintenance Light and standby control On Off Quick programmi...

Страница 36: ...2 11 230Vac RS485 data I O Aux Switch 12 24Vac 7 8 DI2 1 T 3 Aux 22 13 14 16 21 20 19 18 17 N L 15 15 5 A Maximum total current 15A 7 2 A 7 2 A 7 2 A 7 2 A 23 24 AUX2 AUX1 T 2 T 1 Door Switch Door AR2...

Страница 37: ...C Probe T1 probe T2 NO YES Probe T2 enabling evaporator OS2 12 5 12 5 C Probe T2 offset T3 NON DSP CND 2EU Auxiliary probe T3 operation NON probe T3 not fitted DSP temperature T3 to be displayed CND c...

Страница 38: ...ated tables Applications Control Range 50 120 C 55 240 F Resolution 0 1 1 C F Accuracy NTC10K 0 3 C 40 0 70 0 C PTC1000 0 5 C 50 120 C Sensor type selectable NTC10K or PTC1000 Power supply 115 230V 10...

Страница 39: ...A5 A4 A7 A6 A8 L1 N1 N2 N3 N4 N5 N6 L5 L3 L4 L2 L6 L7 1 2 115 230Vac RS485 data I O Aux Switch T3 Aux N L 16 5 A Maximum total current 30A 16 4 A 7 2 A 7 2 A 7 2 A AUX2 AUX1 T2 T1 Door Switch Door AD...

Страница 40: ...Temperature sensor selection With INP SN4 the probes must be the LAE models SN4 with INP ST1 the probes must be the LAE models ST1 OS1 12 5 12 5 C Probe T1 offset T2 NO YES Probe T2 enabling evaporato...

Страница 41: ...d transports HT and LT cold storage rooms plug in cabinets display cases open counters Applications All models come with an alarm buzzer All models are fitted with detachable screw terminals On reques...

Страница 42: ...AUX 7 2 A 7 2 A 12 5 A 7 2 A 5 6 115 230Vac RS485 data I O 7 8 AUX AUXswitch 12 24Vac 17 18 DI2 24 T3 Aux T2 T1 Door switch Door 1 19 21 20 22 23 2 3 4 AH1 5C24W BG...

Страница 43: ...ays off 0 1 the relay contacts follow the on standby state of controller LGT output enabled for light control 2CU output programmed for the control of an auxiliary compressor 2EU output enabled for th...

Страница 44: ...ostat Evaporator Functions Temperature inputs DI1 digital input Power supply Uscite Digital inputs Outputs DI2 digital input Thermostat Auxiliary 1 Auxiliary 2 TTL 230Vac RS485 Serial port All models...

Страница 45: ...Power Supply L N 16 4 A 20 8 A Max total current 30A A B 7 2 A remote AUX 1 DI2 DI1 AUX 2 T1 T2 RS485 PC comm LCD 5S data I 0 BIT25B1S3E B...

Страница 46: ...ut enabled for light control 0 1 the relay contacts follow the on standby state of controller ALO contacts open when an alarm condition occurs ALC contacts make when an alarm condition occurs OA2 See...

Страница 47: ...ications Range 50 120 C 55 240 F Resolution 0 1 1 C F Precision NTC10K 0 3 C 40 0 70 0 C PTC1000 0 5 C 50 120 C Sensor type selectable NTC10K o PTC1000 Power supply 230Vac 10 50 60Hz 3W Technical Data...

Страница 48: ...8 6 1 2 26 29 31 32 33 28 30 27 DI2 Maximum total current 17 5A Power Supply L N data I O RS485 PC comm LCD 5S 16 4 A 16 8 A A B 8 2 A 7A Evaporator fan Condenser fan T3 T2 T1 remote Door 7A AUX 2 AU...

Страница 49: ...Applications Upright refrigerators plug in and supermarket display cases cold stores control panels...

Страница 50: ...ut in Both compressors are turned off at the same time INP SN4 ST1 Temperature sensor selection With INP SN4 the probes must be the LAE models SN4 with INP ST1 the probes must be the LAE models ST1 OS...

Страница 51: ...efrost starts are needed Applications Range 50 120 C 55 240 F Resolution 0 1 1 C F Precision NTC10K 0 3 C 40 0 70 0 C PTC1000 0 5 C 50 120 C Sensor type selectable NTC10K or PTC1000 Rechargeable batte...

Страница 52: ...Power Supply L N data I O RS485 PC comm LCD 5S or RU33 16 4 A 16 8 A A B 8 2 A 7A Evaporator fan Condenser fan T3 T2 T1 remote Door 7A AUX 2 AUX 1 Aux 14 15 16 17 18 19 9 10 11 7 8 6 1 2 26 29 31 32...

Страница 53: ...Applications Cold stores control panels upright refrigerators plug in and supermarket display cases and all those plants where real time defrost starts are needed...

Страница 54: ...tion occurs OA2 See OA1 AUX2 output operation See OA1 2CD 0 120 sec Auxiliary compressor start delay If OAx 2CU the auxiliary output is switched on with a delay of 2CD seconds after the main compresso...

Страница 55: ...to gas used Storage of the latest nine alarms Automatic maintenance management 115Vac or 230Vac power supply by means of built in transformer Connections on screw terminals or quick connectors DIN Rai...

Страница 56: ...I5 DI4 DI1 DI2 2 T NTC10K 1 P Power Supply L N data I O RS485 PC comm 5 1 A 5 1 A A B 5 1 A 5 1 A 7 2 A 0 100 ALR PWM OUT4 OUT3 OUT2 OUT1 I P 0 4 20mA GND 12Vdc 24 5 4 21 20 19 18 9 8 22 11 12 23 14 1...

Страница 57: ...OFF state of output by pressing button While holding down button press button to display the hours of operation multiplied by 1000 the h LED blinks While holding down button press button to display t...

Страница 58: ...PH Alternate Setpoint Pressure refence point is 2SP if DI1 DI2 2SP and the corresponding input is active DBL 10 0 0 0bar Lower neutral zone The state of outputs remains unchanged as long as pressure i...

Страница 59: ...eypad lock option Control of duration of industrial processes control of dough retarders control of cooking time in ovens Applications Outputs Out 16 4 A 240V Power supply 230Vac 10 3W Front protectio...

Страница 60: ...OUT1 START 16 4 A 4 5 6 230V 1 2 10 11 TMR15E A...

Страница 61: ...Applications Control of duration of industrial processes control of dough retarders control of cooking time in ovens...

Страница 62: ...emendo In caso di interruzione di alimentazione il timer si ripropone sempre nello stato di STOP MODO MANUALE Premendo il tasto per 3 secondi si attiva il modo manuale con i tasti e si sceglie lo stat...

Страница 63: ...WBS 01Web Server Wireless Plant Monitoring Monitoring Logging and Programming Software TAB...

Страница 64: ...lish German Italian Spanish Polish etc Minimum System Requirements TAB Available as full optional as described above but also in a low cost version for data logging only This version is called TABLV C...

Страница 65: ...Supervision of the refrigeration process in supermarkets convenience stores shops petrol stations large kitchens food factories cruise ships etc Applications...

Страница 66: ...d difficult both for new and existing units The SWB modules developed by LAE electronics allow all the LAE controllers fitted with a TTL or RS485 port to be Wireless Plant Monitoring incorporated into...

Страница 67: ...nce and intrusions of any type are possible and therefore data reliability and integrity are ensured Safe and reliable The SWB modules may be used to create a fully wireless network a module for each...

Страница 68: ...The TAB software incorporates an easy to use network configuration and management tool...

Страница 69: ...templates for various controllers that LAE makes available and thanks to the automatic detection of the connected devices WBS 01 Web Server In the event that an Internet access is not available remote...

Страница 70: ...tection of the connected devices Communication ability in absence of Internet line and SMS sending by connecting an external analog or GSM GPRS modem Ability to get access to data through Internet por...

Страница 71: ...nnector or points Protection IP67 ST1K C P Sensor type Capacitive Output signal 0 1Vdc Range 0 100 r H Accuracy 5 r H 25 75 r H Sheath 14x40mm Protection IP65 electronics Operating temperature 0 75 C...

Страница 72: ...e 3 wires x 0 24mm2 L 100cm fiber glass points Protection IP65 SPTO Sensor type J thermocouple Range 0 450 C Accuracy 2 5 C or 0 75 in the worst case scenario Response time 10 seconds in water Sheath...

Отзывы: