H50/150 Smart
–
User Manual
Page 21 of 32
Touch Controller
Main Interface
a) Heating output indicator: the icon is switched ON when the unit is heating.
b) Cooling output indicator: the icon is switched ON when the unit is cooling.
c) Real time measurements: process parameters are shown on real time. Liquid InLet and Flowrate parameters are
options and can be removed from the display from FACTORY SETTING
d) Date and time: date and local time. Enter the SYSTEM SETTING window menu to modify them.
e) Running/Alarm status: shows the unit status between Running and Alarm. Available alarms: Temperature Over-
range > Liquid Level > Gas Pressure > Liquid Pressure > Flow Rate >Self Testing > High Temperature > Low
Temperature. Running mode indicates the unit is working properly.
f) Running time: partial running time from the unit switch ON. When the unit is switched OFF the timer is reset.
Maximum time: 10.000 hours.
g) Menu: enter the main MENU
h) Set temperature shows the temperature setpoint value. Touch it to show the TEMPERATURE SETTING window.
a) Heating Output Value
c) Settable
Function
c) Measurements
c)Settable
Function
b) Cooling Output Value
d) Date and Time
g) Menu
f) Run Time
e) Running/Alarm status
h) Set temperature