User Manual LabModul 240
Page 5
www.labtec.de
6. component description
Test stand housing (A)
Thanks to its ergonomic design, the test stand
can be operated fatigue
-
free even during con-
tinuous operation. Operation in computer mo-
de is via PC keyboard and mouse or manually
via the optionally available front panel.
Front panel (B) (optional)
The front panel contains the control buttons
and a multi
-
function LCD display showing print
and system information.
The following abbreviation is used:
LP = Low pressure (
-
50...+50 mbar)
MP = medium pressure (0...25 bar)
Multifunction probe (C)
The test head is a fixed head that can be infla-
ted by an electric pump so that full face masks
of any make can be sealed without problems.
A measuring point is installed in the test head
which measures the internal pressure in the
mask or lung automat.
Medium pressure device connections (D, E)
Plug nipple for connecting the medium pres-
sure line of the compressed air breathing ap-
paratus pressure regulator to be tested.
Quick coupling for connecting the medium
pressure line of the regulator to be tested.
CPS test connection and system vent (F)
Quick coupling for filling CPS. For all other test
procedures, the coupling must be connected
to the supplied plug
-
in nipple/silencer for sys-
tem venting.
Interface (H)
The USB interface for data transfer to a PC is
located on the left side of the housing
(LabModul 240 MT has no function with the
manual device).
Vacuum pump
With the built
-
in vacuum pump, the required air
currents are generated and the probe is infla-
ted or deflated. The desired air flow rate can
be set in the software from 2...12 l/min.
Processor
The built
-
in microprocessor takes over the
control and signal conversion.
7. description of the operating elements
1
2
LCD multi function display
Digital display of status of operation
Display for low pressure (LP)
Display for medium pressure (MP)
3
Push button for test head inflation
Pressing causes the test head to be inflated out
of the test head
4
Push button for
test head aspiration
Pressing causes the air to be aspirated out of
the test head
5
Push button to generate positive pressure
Forcing pump +5
l/min is set in motion by press-
ing this button
6
Push button to generate negative pressure
Forcing pump –
5
l/min is set in motion by press-
ing this button
7
Push button for low pressure decompress-
ing
Pressing causes the air to be aspirated out of
the test equipment
8
Switch button for forcing pump
Forcing pump +10
l/min is set in motion by
pressing this button. LED shows function.
Pressing button again switches it off.
9
Switch button for forcing pump
Forcing pump
-
10
l/min is set in motion by
pressing this button. LED shows function.
Pressing button again switches it off.
10
Switch button for CPS inflation
Pressing causes the CPS to be inflated out of
the CPS, Medium pressure is needed in the test
bench.
11
Switch button for CPS deflation
Pressing causes the air to be aspirated out of
the CPS
12
Push button for medium pressure decom-
pression
Pressing causes the air to be aspirated out of
the test equipment