User Manual Labtec ECO

 

page 9

 

www.labtec.de

 

Step

 

Activity

 

Comment

 

operating 

elements

 

1.1.1

 

Close breathing connection to mask with sealing 

plug.

 

If necessary, moisten the exhalation 

valve of the mask. 

 

 

1.1.2

 

Place the mask on the test head and tighten the 

mask strap. Fill test head with head inflation 

button

 

The inflation time varies from mask 

type to mask type.

 

    

 

1.1.3

 

Set test vacuum with negative pressure button 

to setpoint value 

-

 # mbar. Switch on stop watch.

 

Adjust the test pressure with the low 

pressure relief button if necessary.

 

 

1.1.4

 

Read the pressure difference after # min. on the 

display (LP) and compare with setpoint.

 

 

 

1.1.5

 

Deflate the test head with button for test head 

deflation. Remove mask from test head.

 

 

 

 

 

 

 

1.2.1

 

Seal the exhalation valve of the mask with the 

adapter. Close breathing connection to mask 

with sealing plug.

 

Test only possible with tightly fitted 

exhalation valve. Use the manufac-

turer's specific adapters.

 

 

1.2.2

 

Place the mask on the test head and tighten the 

mask strap. Fill test head with head inflation 

button

 

The inflation time varies from mask 

type to mask type.

 

 

1.2.3

 

Set test vacuum with negative pressure button 

to setpoint value 

-

 # mbar. Switch on stop watch.

 

Adjust the test pressure with the low 

pressure relief button if necessary.

 

 

1.2.4

 

Read the pressure difference after # min. on the 

display (LP) and compare with setpoint.

 

 

 

1.2.5

 

Deflate the test head with button for test head 

deflation. Remove mask from test head.

 

 

 

 

 

 

 

1.3.1

 

Close breathing connection to mask with sealing 

plug.

 

 

 

1.3.2

 

Place the mask on the test head and tighten the 

mask strap. Fill test head with head inflation 

button

 

The inflation time varies from mask 

type to mask type.

 

 

1.3.3

 

Switch on vacuum pump with switch 10l/min 

Exhalation to produce constant flow.

 

 

 

1.3.4

 

Read the opening pressure on the display (LP) 

and compare it with the setpoint.

 

 

 

1.3.5

 

Switch off the pump by pressing the 10/lmin 

Exhalation switch again.

 

 

 

1.3.6

 

Deflate the test head with button for test head 

deflation. Remove mask from test head.

 

 

 

A 1 Testing of full face breathing masks

 

1. 

 

Complete device (full face mask)

 

1.1 

 

Leak test with open exhalation valve at 

-

 # mbar 

   

1.2 

 

Tightness test with closed exhalation valve at + # mbar

 

1.3 

 

Opening pressure of the exhalation valve

 

Содержание ECO static

Страница 1: ...ec de User Manual Labtec ECO static dynamic Labtec No Content Page 1 foreword 2 2 safety instructions 2 3 device description 3 4 technical data 4 5 component description 5 6 operating elements 6 7 com...

Страница 2: ...ained and certified Labtec service technician and tested and calibrated by him or her using a Labtec calibration device The list of current certified employees trai ning partners and service technicia...

Страница 3: ...PC connec ted to test device via swivel arm mounting incl pre installed LabNet Profi test software Module S safety valve test System com ponent and module for automatic medium pressure increase for SC...

Страница 4: ...on bar 0 25 0 5 low pressure sensor measuring range precision mbar 50 0 50 0 05 complete device Max permissible operating pressure bar 400 power supply V Hz 230 240 50 60 total power electricity W max...

Страница 5: ...linder or pipe supply Attention Never connect or disconnect lines under pressure Device connection E High pressure device socket for connecting the pressure regulator of the SCBA to be tested CPS test...

Страница 6: ...switches it off Switch button for 10 l m inhalation Forcing pump 10 l min is set in motion by pressing this button LED shows function Pressing button again switches it off Switch button for CPS infla...

Страница 7: ...atus is screwed completely into the inter nal thread of the device connection For tests in the medium pressure range dis connect the medium pressure line of the SCBA at the Euro coupling and connect i...

Страница 8: ...C Guideline 2016 42 EG and EMC 2014 30 EU including all changes and addendums to date thereto The above mentioned product meets the following standards and guidelines Nous Vous Confirmons que l appare...

Страница 9: ...h head inflation button The inflation time varies from mask type to mask type 1 2 3 Set test vacuum with negative pressure button to setpoint value mbar Switch on stop watch Adjust the test pressure w...

Страница 10: ...r a preset time min read the pressure change on the display LP and compare with the setpoint 2 1 5 Completely vent the test device All further tests require high pressure Check correct device connecti...

Страница 11: ...ith main valve switch 2 2 7 Completely vent the test device 2 4 1 Connect demand valve in combination with a SCBA base unit to the test device Check correct device connection 2 4 2 Open high pressure...

Страница 12: ...t pressure until PA pressure gauge shows the set point Read off the SCBA manometer 3 1 5 Read the high pressure on the display HP and compare it with instrument pressure opt Repeat test steps 3 1 4 an...

Страница 13: ...test device 5 3 1 Connect demand valve in combination with a SCBA base unit to the test device Check correct device connection 5 2 1 Open high pressure main valve with main valve switch High pressure...

Страница 14: ...ation with a SCBA base unit to the test device Check correct device connection 5 4 2 Open high pressure main valve with main valve switch High pressure flows into the unit 5 4 3 Close high pressure ma...

Страница 15: ...des Schalters CSA f llen das Bef llen des Anzugs stoppen when the required tightening internal pressure is reached 7 1 8 Press the CPS inflation switch again to stop fil ling the suit Setting the requ...

Отзывы: