7  Trouble shooting 

 

   

 

 

’    

 

 

   

 

The performance of the microscope can t be made fully because of

unfamiliar using

 

 

 

 

 

.

This table will give some advices

 

Trouble

Cause

Remedy

1.

 

Double images

 

   

Interpipillary distance is

 

not correct

 

Readjust it

 

   

Diopter djustment is not

 

correct

 

Readjust it

   

 

Magnification of each

   

 

 

eyepiece is not the

 

 

same size

 

 

 

Mount the same size

eyepiece

2.

 

   

 

Dirt appears in the

   

field of view

   

 

Dirt on the specimen

 

 

 

Clean the specimen

   

 

   

Dirt on the surface of

eyepiece

 

 

 

Clean the surface

3.

   

 

   

Image is not clear

   

 

   

Dirt on the surface of

 

 

the objective

 

 

 

Clean the objectives

4.

   

 

 

Image is not clear

 

 

 

while the focus

changing

 

   

Diopter adjustment is

 

not correct

 

 

Readjust the diopter

   

 

Focus is not correct

 

 

Readjust the focus

5.

 

 

 

The focusing knob is

 

not smooth

 

 

   

The focusing knob is

 

 

too tight

       

Loosen it to a

 

 

suitable position

6.

 

   

The image is obscure

   

 

 

because of the head

 

   

slipping down by itself

 

during observation

 

 

   

The focusing knob is

 

too loose

       

Tighten it to a

 

 

suitable position

7.

 

 

Incision image

   

 

 

appears in the field of

     

 

 

view or of the video

view

 

   

   

The pole is not in

 

correct position

   

     

Pull or push it to

 

 

the correct position

8.

 

 

 

Eyes fell tired easily

 

   

Diopter adjustment is

 

not correct

 

 

Adjust the diopter

   

   

Brightness of light is not

correct

 

 

Adjust the brightness

9.

 

 

 

 

Bulb does not work

 

 

   

when the switch is on

 

 

No power in

 

 

Check the

 

 

 

connection with the

 

power supply

 

 

 

 

 

The bulb was not insert

 

correct

   

Insert it correctly

   

Bulb is wrong

 

   

 

Replace with a new

one

10.

   

 

 

Bulb is burned out

suddenly

 

 

 

Use the wrong bulb

 

   

Replace with a

 

correct one

 

   

 

The voltage is too high

 

 

Control the voltage

 :

 

 

Eg use voltage

regulator

11.

   

 

Brightness is not

enough

 

 

 

Use the wrong bulb

 

   

Replace with a

 

correct one

 

   

 

The voltage is too low

 

 

 

Increase the input

voltage

15

STEREO ZOOM MICROSCOPE

www.labstac.com

Содержание MSC41-060ST

Страница 1: ...OPERATION MANUAL STEREO ZOOM MICROSCOPE MSC41 060ST...

Страница 2: ...us 4 5 9 Adjust the interpupillary distance 4 6 9 Use Eyepiece shields 4 8 9 Install the illumination device 4 9 10 Choose the optical system 4 10 10 Mount the photo eyepiece and the PK mount adapter...

Страница 3: ...s must be kept clean Fine dust on the surface of the lens should be blown off with hand blower or wiped off gently with a soft lens tissue Fingerprints or oil marked on it should be wiped off with a t...

Страница 4: ...tac com 4 STEREO ZOOM MICROSCOPE Head Eyepiece shield Eyepiece Diopter adjust ring Lock screw Glass stage Clip Bracket Focusing knob Illumination device Base Fuse Bottom variable light control Bottom...

Страница 5: ...ww labstac com Base Bracket Photocube Illumination device Focusing knob device Eyepiece shield Eyepiece Diopter adjust ring Glass stage Head Lock screw Clip Top variable light control Bottom variable...

Страница 6: ...3 Assemblage www labstac com 6 STEREO ZOOM MICROSCOPE...

Страница 7: ...PK mount adapter Photo eyepiece Eyepiece shield C mount Eyepiece Illumination device Head Clip Glass stage Base 7 STEREO ZOOM MICROSCOPE www labstac com...

Страница 8: ...is tight otherwise is loose 2 The suitable position of the tightness can make the adjustment more comfortable and prevent the focusing bracket from slipping down by its weight during the observation 2...

Страница 9: ...Turn and remove the mounting ring from the eyepiece 5 Fig 2 Clean the eyepiece micrometer and mount it to the mounting ring with the inscription side downward 3 Gently twist the mounting ring with the...

Страница 10: ...light 1 Turn the adjustable light knob according to the sign marked on the base along the clockwise the brightness will 9 be added otherwise it will be weakened Fig 4 12 Fig 11 4 12 Replace the lamps...

Страница 11: ...1 100 X mm 2 100 X mm 3 100 X mm 4 100 X mm Binocular 60 ST Tri ocular 60 ST TR Big conversion lens 0 5 165 SZMAO mm 1 5 45 SZMAO m m 2 30 SZMAO m m Focusing arm 1 SZMA Stand 1 SZMST 2 SZMST 3 SZMST 1...

Страница 12: ...www labstac com 12 STEREO ZOOM MICROSCOPE...

Страница 13: ...fixed regardless of the magnification factor Total mag Objective mag X Auxiliary mag X Auxiliary mag Eyepiece field Diameter of field of view mm Objective mag X Auxiliary objective mag Photo adaptor m...

Страница 14: ...eyepiece tube 54 75 Range of single adjustment mm Diopier adjustment Range of single adjustment 5D Mount with auxiliary objectives Screw hole 48 0 75 M Objecti ve Field of view 20mm Main body Mount th...

Страница 15: ...Loosen it to a suitable position 6 The image is obscure because of the head slipping down by itself during observation The focusing knob is too loose Tighten it to a suitable position 7 Incision image...

Страница 16: ...12 The bulb flickers or the brightness is unstable The bulb will burn out soon Replace with a new one The bulb was not inserted correctly Insert it correctly www labstac com 16 STEREO ZOOM MICROSCOPE...

Страница 17: ...Email contact labstac com Website www labstac com 17 STEREO ZOOM MICROSCOPE www labstac com...

Отзывы: