SERVICE AND ANALYSIS 

REFRIGERATION 

CHART 

SYSTEM 

MALFUNCTION 

POSSIBLE CAUSE 

SOLUTION 

Compressor will not start - no hum 

1. Line cord not plugged in. 

1. Plug in the cord. 

2. Fuse removed or blown. 

2. Replace fuse. 

3. Overload protector tripped. 

3. Refer to electrical section. 

4. Temp control stuck in open position. 

4. Repair or replace temp control. 

5. Wiring improper or loose. 

5. Check wiring against diagram. 

Compressor will not start - hums but 

1. Low voltage to unit 

1. Determine reason and correct 

trips on overload protector. 

2. Relay failing to close 

2. Determine reason and correct, replace 

if necessary 

3. Starting capacitor defective. 

3. Determine reason and replace 

4. Improperly wired. 

4. Check wiring against diagram 

Compressor starts but does not switch 

1. Low voltage to unit. 

1. Determine reason and correct. 

off of start winding. 

2. Relay failing to open. 

2. Determine reason and correct, replace if 

necessary. 

3. Run capacitor defective 

3. Determine reason and replace. 

4. Comp Mt winding is open or shorted 

*4. Determine cause. correct, and replace comp 

Compressor starts and runs, but 

1. Additional current passing through 

1. Check wiring diagram. Check for added fan 

short cycles on overload protector. 

overload protector. 

motors, pumps, etc. connected to wrong 

side of protector. 

2. Low voltage to unit. 

2. Determine reason and correct. 

3. Overload protector defective. 

3. Check current, replace protector. 

4. Run capacitor defective. 

4. Determine reason and replace. 

5. Excessive discharge pressure. 

*5. Check ventilation, restrictions in cooling 

medium, restrictions in refrigeration system. 

6. Compressor too hot - return gas hot. 

*6. Check refrigerant charge (fix leak if 

       necessary). Check air flow across condenser. 

Unit runs OK, but short cycles. 

1. Overload protector. 

1. Check wiring diag for correct wiring 

2. Cold control. 

2. Differential set too close. 

3. Overcharge. 

*3. Reduce refrigerant charge. 

4. Air in system. 

*4. Recover and recharge. 

5. Undercharge. 

*5. Fix leak and recharge with refrigerant. 

Unit operates long or continuously. 

1. Dirty condenser 

*1. Clean condenser 

2. Shortage of refrigerant. 

2. Fix leak, add charge, correct charge 

3. Temp control contacts stuck or frozen 

3. Replace Temp control 

4. Evaporator coil iced. 

*4. Defrost 

5. Restriction in refrigeration system. 

5. Determine location and remove. 

Start capacitor open, shorted or blown. 

1. Relay contacts not opening properly. 

1. Replace relay 

2. Low voltage to unit. 

2. Determine reason and correct. 

3. Improper relay. 

3. Replace. 

Run capacitor open, shorted or blown. 

1. Improper capacitor. 

1. Determine correct size and replace. 

2. Excessively high line voltage 

2. Determine reason and correct. 

(110% of rated max). 

Relay defective or burned out. 

1. Incorrect relay. 

1. Check and replace. 

2. Line voltage too high or too low. 

2. Determine reason and replace. 

3. Relay being influenced by loose 

3. Remount rigidly. 

vibrating mounting. 

Space temperature too high. 

1. Control setting too high. 

1. Reset control. 

2. Overcharged with refrigerant. 

*2. Recover refrigerant and recharge with 

proper charge specified on dataplate. 

3. Inadequate air circulation. 

3. Improve air movement. 

Cooler freezing beverage. 

1. Temperature control 

1. Reset control. 

Unit noisy. 

1. Loose parts or mountings. 

1. Find and tighten. 

2. Tubing rattle. 

2. Reform to be free of contact. 

3. Bent fan blade causing vibration. 

3. Replace blade. 

4. Fan motor bearings worn. 

4. Replace motor. 

LAB

2QH&RPSDQ\‡0DQ\6ROXWLRQV

RepCo

101 Witmer Road, Suite 700 • Horsham, PA  19044

Phone  1-800-521-0754 • Fax  215-442-9202

www.labrepco.com • [email protected]

Содержание LABL-27-HG

Страница 1: ...n includes complete refrigeration System cabinet construction finish IMPORTANT PLEASE RETAIN FOR YOUR RECORDS LABL 27 HG LABL 38 SG LABL 45 SG LABL 49 HG AND LABL 74 HG 101 Witmer Road Suite 700 Horsh...

Страница 2: ...ustable to suit requirements See Instruction Sheet ILA 0904 for installing e Door Removal and Adjustment 1 Slide Doors Each door has its own Closing Spring located at the top and is positioned for Pro...

Страница 3: ...r flow over the pan hastens condensate evaporation so that external drain plumbing is not required 3 Maintenance Caution Disconnect Electrical service before accessing mechanical enclosures having mov...

Страница 4: ......

Страница 5: ......

Страница 6: ...LL SCALE SHELF SUPPORT FRONT WITH 1 4 TAB FULL SCALE NOTES INSTRUCTIONS 1 PLACE SHELF SUPPORTS WITH 1 4 TABS INTO PILASTER AT DESIRED POSITION NOTE THAT SUPPORTS WITH 1 4 TABS ARE LOCATED AT FRONT ONL...

Страница 7: ...hields by using the identification mark on the end of the bulb Properly installed bulbs will have the mark centered between the halves of the lamp holder On units with shields the identification mark...

Страница 8: ......

Страница 9: ...sy Off Oven Cleaner Apply generous coating Excellent removal fatty acids miikstone Allow to stand for 10 to 15 Satisfactory for use on all where swabbing or rubbing minutes Rinse Repeated finishes is...

Страница 10: ...if necessary Check air flow across condenser Unit runs OK but short cycles 1 Overload protector 1 Check wiring diag for correct wiring 2 Cold control 2 Differential set too close 3 Overcharge 3 Reduce...