LABORIE Urocap IV Light Service Manual UCL-SM01 V7.00
ELECTROMAGNETIC COMPATIBILITY (EMC)
Standards to which conformity is declared:
ANSI C63.4:2009
Methods of Measurement of Radio-Noise Emissions from Low-
Voltage Electrical and Electronic Equipment in the Range of 9 kHz to
40 GHz.
IEC 60601-1-2:2007
(Ed3.0)
Medical Electrical Equipment
–
Part 1-2: General Requirements for
Safety
–
Collateral Standard: Electromagnetic Compatibility.
CISPR 11:2010
Industrial, Scientific and Medical (ISM) Radio-Frequency
Equipment
–
Electromagnetic Disturbance Characteristics
–
Limits and Methods of Measurement.
CISPR 22:2008
Information Technology Equipment - Radio Disturbance
Characteristics - Limits and Methods of Measurement.
IEC 61000-3-2:2009
Limits for Harmonic Current Emissions (equipment input current
= 16A per phase).
IEC 61000-3-3:2008
Limitation of Voltage Changes, Voltage Fluctuations and Flicker in
Public Low-Voltage Supply Systems, for equipment with rated current
= 16A per phase and not subject to conditional connection.
IEC 61000-4-2:2008
Testing and Measurement Techniques - Electrostatic Discharge
Immunity Test.
IEC 61000-4-3:2010
Testing and Measurement Techniques -
Radiated, Radio-Frequency, Electromagnetic Field Immunity Test.
IEC 61000-4-4:2012
Testing and Measurement Techniques -
Electrical Fast Transient/Burst Immunity Test.
IEC 61000-4-5:2005
Testing and Measurement Techniques - Surge Immunity Test. IEC
61000-4-6:2006
Testing and Measurement Techniques - Immunity to Conducted
Disturbances, Induced by Radio-Frequency Fields.
IEC 61000-4-8:2009
Testing and Measurement Techniques - Power Frequency
Magnetic Field Immunity Test.
IEC 61000-4-11:2004
Testing and Measurement Techniques - Voltage Dips, Short
Interruptions and Voltage Variations Immunity Tests.
ISO 17025:2005
General requirements for the competence of testing and
calibration laboratories.