LABORIE Urocap IV Light Service Manual UCL-SM01 V7.00
Table 5: Troubleshooting - Urocap IV
S
YMPTOM
(
S
)
P
OSSIBLE
C
AUSE
(
S
)
C
HECK
/ C
ORRECTIVE
A
CTION
(
S
)
Uroflowmeter
disconnected
while test in
progress?
Battery power
low or
completely
drained?
1.
Tap
Stop
.
2.
Plug in the disconnected device’s
battery charger and plug the charger
into an electrical outlet. Make sure
connections are secure.
3.
Tap the device icon with the
Disconnected status, and wait for the
reconnection to the device.
The test data acquired before the
battery failure can be printed by
tapping the Reprint Last Test icon.
Uroflowmeter
signal shows
vibration and/or
spike patterns?
Plastic beaker is
touching the flow
funnel?
Reposition the Uroflowmeter and check
again.
Patient has
touched the
Uroflowmeter
with their feet?
Ask patient to remain calm during procedure.
Floor is
unsteady?
Move to a more solid foundation.
Incorrect flow or
volume readings?
Beaker not
properly seated
on Uroflowmeter
dish?
Adjust beaker position.
Funnel is touching
beaker?
Adjust commode chair or Urocap IV.
Incorrect beaker
in use?
Use beakers supplied by LABORIE ONLY.
Does the
Uroflowmeter
require
calibration?
Calibrate the Urocap IV and start over.