LABORIE Urocap III Owner’s Manual
UC3-UM01
61
IEC 60601-1-2: 2014 – TABLE 6 REQUIREMENTS
Recommended separation distances between portable and mobile RF communications equipment and the
Urocap III
®
.
The
Urocap III
®
is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the
Urocap III
®
can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the
Urocap III
®
as recommended below,
according to the maximum output power of the communications equipment.
WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than
30 cm (12 inches) to any part of the Aquarius system, including cables specified by the manufacturer. Otherwise, degradation of the performance of this
equipment could result.
Rated maximum
output power
of transmitter
(W)
Separation distance according to frequency of transmitter
(m)
150 kHz to 80 MHz
𝑑𝑑
= 1.17
�𝑝𝑝
80 MHz to 800
MHz
𝑑𝑑
= 1.17
�𝑝𝑝
800 MHz to 2.5
GHz
𝑑𝑑
= 2.33
�𝑝𝑝
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.70
3.70
7.37
100
11.70
11.70
23.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in metres (m) can
be estimated using the equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating
of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.