It is always possible to connect the Bluetooth flowmeter to the Solar Blue
module by using a digital MMS-bus cable. All of the following three conditions
must be met:
•
The Solar Blue module needs to communicate with the computer via USB.
•
The Solar Blue module must be powered by the external power supply.
•
The flowsensor must be connected to one of the digital MMS-bus ports of
the Solar Blue module. (Note: it is not possible to connect the power
adapter to the flowsensor as it is powered by the Solar Blue external power
supply.)
Status LED flowsensor
The following table describes what the LED colors and status indicate:
LED
Indication
Orange on
IPL error. Please contact your MMS representative.
Green on
Digital cable is connected.
Green blinking
(short periods off)
Bluetooth communication flowsensor powered by power
adapter.
Green blinking
(long periods off)
Bluetooth communication flowsensor powered by
batteries.
Red blinking
Low battery message, place new batteries in the battery
compartment.
Off
No or empty batteries. Place new batteries in the battery
compartment.
Power adapter (optional) is not connected. Reconnect
the power adapter.
Bluetooth flowsensor in low power mode.
Bluetooth connection is lost. In case you are using the
Bluetooth dongle, check whether the dongle is missing
or broken.